戻る
「早戻しボタン」を押すと検索画面に戻ります。 [閉じる]

コーパス検索結果 (1語後でソート)

通し番号をクリックするとPubMedの該当ページを表示します
1 al of natural Si isotopic composition as the calibrant.
2 sing the reference standard (RS) material as calibrant.
3 is a strong candidate as a negative ion mode calibrant.
4 was due to loss of HIV DNA from the 8E5 cell calibrant.
5  application, because of its dependence on a calibrant.
6 ans of ratio counting using a colloidal gold calibrant.
7 nds, that is, the compound of interest and a calibrant.
8  has been limited by the lack of appropriate calibrants.
9 lies on mass differences instead of internal calibrants.
10 milar chemical properties between lipids and calibrants.
11 ix, indicating no need to use matrix-matched calibrants.
12 ion manipulations (SLIM) using three sets of calibrants.
13 sed as suitable for value assigning such DNA calibrants.
14 108% for all cereals matrices using external calibrants.
15 of labeled metabolites using a mixture of RT calibrants.
16 sociated with pRSETMON-02 as opposed to gDNA calibrants.
17 tter alternative to conventional seed powder calibrants.
18 eptides to the isotopically labeled internal calibrants.
19 the detected species that serve as effective calibrants.
20 ge biomolecules that can then be used as CCS calibrants.
21 minal residues and provide valuable internal calibrants.
22 accuracy MS and MS/MS spectra using separate calibrant and analyte regions on the target plate, avoid
23 ismatching while the same molecular class of calibrant and analyte was maintained yielded larger erro
24 ass standard without the necessity of mixing calibrant and analyte.
25 fication of the 50 compounds using a generic calibrant and mass calibration achieved an average error
26 pass calibration methods, which require both calibrants and analytes to experience the same number of
27             Stable isotope-labeled surrogate calibrants and internal standards were used to achieve m
28 N(2)) and Omega(He) values for the drug-like calibrants and two diastereomers were measured using a d
29 as well as the laborious nature of requiring calibrants and unknown compounds to be subjected to iden
30 -MS instrumentation relies on the use of CCS calibrants, and here, we present data from a family of p
31 ccurate calibration with a minimal number of calibrants, and the ability to extract unknown parameter
32 re analyzed at the same TW conditions as the calibrants, and we observed anomer splitting at pathleng
33 n CCS values, which can impart biases if the calibrants are not structurally similar to the analytes.
34                                        These calibrants are often internal and behave in a fashion si
35                                      Current calibrants are sometimes not designed to satisfy these r
36                                         Such calibrants are suitable to determine the reduced mobilit
37 neration of bespoke fluorescent protein (FP) calibrants, assays to determine protein concentration an
38 ration levels were analyzed together with 41 calibrants at 6 known concentrations by the laboratories
39         In this paper, we introduce a new IM calibrant (CCS and K0) using singly charged sodium adduc
40 ification of target concentration in primary calibrant certified reference materials (CRMs), both lam
41 for suitable ion mobility spectrometry (IMS) calibrant compounds is ongoing and necessitates the use
42 age of the methods presented here is that no calibrant compounds need to be added to the mobile phase
43 ure component relative to "internal affinity calibrants", compounds of known affinity for the target.
44 area and plotting the area ratio against the calibrant concentrations.
45                                    Since all calibrants consist of the same species and differ only i
46  obtain sufficiently high number of internal calibrants covering the entire ranges of mass-to-charge
47 sing an internal standard curve and need for calibrant diluent, and takes into account the individual
48  100-500 nmol/mol, with NO and NO2 reactants/calibrants diluted down from standards with nominal mole
49 urthermore, high method recoveries indicated calibrants do not need to undergo clean-up to account fo
50                                         pDNA calibrants exhibited better performance characteristic i
51 ce liquid chromatography, is also a suitable calibrant for CCS estimations.
52 his work establishes a new type of inorganic calibrant for IM-MS allowing sizing, structural analysis
53 e suitability of IsoF as a negative ion mode calibrant for IMS applications.
54 ifs, ultimately enabling the use of only one calibrant for these compounds.
55                 However, currently available calibrants for CCS measurement using TWIM are predominan
56 Li, Na, K, Rb, Cs; R = aryl, Et; n = 1-6) as calibrants for electrospray ionization (ESI) TWIMS.
57 es multiple ion isolations of an analyte and calibrants for internal calibration at each stage in a M
58 tudy of large biomolecules, however reliable calibrants for large anions have not been available.
59 cy and reproducibility of PCs and PEs as CCS calibrants for phospholipids against different classes o
60  organic compounds that are commonly used as calibrants for qNMR spectroscopy: benzoic acid, dimethyl
61         Atmospheric moisture can contaminate calibrants for quantitative nuclear magnetic resonance (
62                          However, a suitable calibrant has yet to be accurately analyzed and agreed u
63 stigations for a negative ion detection mode calibrant have been more limited.
64                       Many positive ion mode calibrants have been investigated, whereas investigation
65  injection and trapping of the internal mass calibrant (i.e., poly(ethylene glycol) with a 1000 Da av
66 eveloped calibration curves using common CCS calibrants [i.e., tetra-alkylammonium salts, polyalanine
67  qHNMR using a single, nonidentical internal calibrant (IC).
68 rge effects is to introduce an internal mass calibrant (IMC) using a dual electrospray ionization (ES
69 ose of the healthy hemisphere as an internal calibrant in addition to routinely used external standar
70  show that dextran, already widely used as a calibrant in high performance liquid chromatography, is
71 nriched metalloprotein standard for use as a calibrant in species-specific isotope dilution analysis
72 l(2)O(3)) has been widely used as a pressure calibrant in static high-pressure experiments and as a w
73 ope-labeled acyl-CoAs were used as surrogate calibrants in the same matrix.
74                                         Such calibrants include buffer solutions, "anchor" molecules
75 r phospholipids against different classes of calibrants, including polyalanine (PolyAla), tetraalkyla
76 ins some question regarding how selection of calibrants influences TWIMS CCS measurements.
77 n "unknown" concentration without the use of calibrants, internal standards, or a calibration curve.
78 r analyte quantitation that does not rely on calibrants, internal standards, or calibration curves bu
79 ffects that would result from the spiking of calibrants into the sample.
80              This report presents a study of calibrant ion influence on CCS determination via TWIMS.
81  through the use of an internal energy-input calibrant ion that ensured reproducible energy depositio
82    Furthermore, we investigate the number of calibrant ions as well as their required m/z range in or
83                                 In addition, calibrant ions desorbed from a separate sample spot can
84  a method for the controlled introduction of calibrant ions in highly accurate mass measurements usin
85 well-defined series of positive and negative calibrant ions using either electrospray ionization (ESI
86     Since a single concentration of internal calibrant is necessary to calculate the study sample con
87  when conducting experiments in which a mass calibrant is needed or when multiplexing in a tandem MS
88 ignals using sucrose and ammonium sulfate as calibrants is presented.
89 The experimental results showed that the one-calibrant kinetic calibration technique can be used for
90                      In open bed form, a one-calibrant kinetic calibration technique was accomplished
91  We demonstrate that structurally mismatched calibrants lead to larger errors in calibrated CCS value
92                                          The calibrant-loaded approach can provide both total and fre
93 se and analyte uptake between the sample and calibrant-loaded extraction phase, CL-EP, with a finite-
94  ICP-OES analysis for production of accurate calibrants may increase quantitative accuracy and give g
95 sue sections by MALDI t-wave IM-MS using CCS calibrants measured by MALDI drift cell IM-MS.
96                          Mismatching of both calibrant molecular class and charge state resulted in t
97 tandardization of seminal plasma derived PSA calibrant molecular form mass concentrations and purific
98 centroided m/z values of the known, internal calibrant molecular ions versus that calculated from the
99 ltiomic mixtures, several different types of calibrants must be used to obtain the most accurate CCS
100 e of the mass error relative to the internal calibrant never exceeded 5.06 ppm.
101           This is of particular concern when calibrants of the same molecular class (e.g., carbohydra
102  accomplished by loading benzophenone3-d5 as calibrant on the C18 coating to quantify all nonpolar co
103 lysis is achieved using ICA by incorporating calibrants on a separate spot from the sample of interes
104 his study, we have chosen five potential IMS calibrants on the basis of their rating against seven cr
105 tly hindered by the requirement for external calibrants on which a working pH scale is set.
106 1-min aFT-IMS-MS sweeps without the need for calibrants or correction factors.
107 rom 0.04-2.5microg/mL and stable derivatized calibrants over 7days.
108 ed CCS values while the structurally matched calibrants, PCs and PEs, gave highly accurate and reprod
109 ose are quite unreliable, leading to missing calibrant peaks in spectra and correspondingly lower mas
110                              With the set of calibrants presented here, the m/z range of 40-400 kDa i
111 t to be unsuitable for the purpose of single-calibrant quantification.
112 alculated based on the fact that 100% of the calibrant reaches the plasma through the MDG, whereas fo
113                   We present the kinetics of calibrant release and analyte uptake between the sample
114 endations are made herein for each potential calibrant's suitability as a standard for the wider IMS
115  4 which demonstrate their use as a TW-IM-MS calibrant set to facilitate characterization of very lar
116         The calibration errors for the three calibrant sets (Agilent tuning mixture, polyalanine, and
117                                              Calibrant solutions are delivered from reservoirs fabric
118 ol to scavenge residual hydrogen peroxide in calibrant solutions provided analytically sensitive and
119 ons was established and was used to identify calibrant solutions that can be employed for the simulta
120 th this system, we have been able to deliver calibrant solutions, washing buffers, or "test" solution
121 irectly determined CCS values for 19 protein calibrant species under three solution conditions (yield
122  of mass of standard reference materials and calibrants, such as ZrH(2), 2,5-diiodothiophene, and low
123                          The use of internal calibrants (the so-called lock mass approach) provides m
124 a universally applicable and robust internal calibrant, the tripeptide Asn3.
125                              Using the lipid calibrants, the majority of the CCS values of several cl
126 andard, and therefore, the RS can serve as a calibrant to correct for the difference between instrume
127 o be used with a single external nitrogenous calibrant to quantify nitrogen-containing compounds of k
128 plored the validity of using "normal" charge calibrants to calibrate for charge-reduced proteins and
129 TW IM-MS) instrumentation rely on the use of calibrants to convert the arrival times of ions to colli
130 nable absolute quantification using internal calibrants to determine copies per cell of MHC antigens,
131   These instruments often rely on drift-time calibrants to perform qualitative identification.
132  isotope labeled (SIL) standards as internal calibrants to simultaneously quantify authentic analytes
133                                      The PSA calibrants used to create calibration curves in these as
134 ddressed and the first RI bank based on CDEA calibrants was developed.
135   Parallelism between analytes and surrogate calibrants was systematically verified in plasma across
136 antification of nucleic acid not requiring a calibrant, we demonstrated that inclusion of a carrier n
137 nation of small molecule and polypeptide CCS calibrants, we successfully determined the nitrogen CCS
138 e been corrected through the use of internal calibrants (well-characterized standards introduced with
139                Standard curves of the spiked calibrants were generated with concentrations ranging fr
140                                              Calibrants were made of nitrocellulose membranes doped w
141                               Matrix-matched calibrants were needed to compensate for over recovery o
142                               Two sets of 18 calibrants were selected for each of ESI+ and ESI-based
143          The development of phospholipid CCS calibrants will enable high-accuracy structural studies
144 -tagged peptides can further act as internal calibrants with high confidence in a PMM to identify the
145 -to-charge ratios and abundances of internal calibrants without a need for theoretical description of
146 standard oxidation-reduction potential (ORP) calibrants, ZoBell's and Light's solution.

 
Page Top