コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 the tryptophan fluorescence lifetime (60 mus dead time).
2 ted association reaction in the stopped-flow dead time.
3 higher column outlet pressures with minimal dead time.
4 values in a "burst phase" within the mixing dead time.
7 ssociated with refolding occurred during the dead time (4 ms) of the stopped-flow instrument, suggest
8 ly verify a theory that takes the effects of dead time, afterpulsing, and the finite sampling time on
10 high-speed separations by reducing detector dead time and by shifting optimal carrier gas velocity t
12 increase in detection rate at an acceptable dead time and introduce a new way of tuning the detector
13 ith an assay that minimized the experimental dead time and which allowed for detection of N-acetyltyr
15 that takes nonideal detector effects such as dead-time and afterpulsing into account is developed and
16 mus using a Monte Carlo model simulating the dead-time and background effects in the standard neutron
17 information up to the start of the saccadic dead time, and (3) variability in saccade latency does n
18 d improvements in pulse shaping, integration dead time, and triggering, has an improved count-rate ca
22 onditional reset of clock qubits and minimal dead time between repetitions by implementing ancilla-ba
24 d that the population formed during the 4 ms dead time contained multiple species that are stabilized
25 the application of several corrections, with dead-time correction and background subtraction being pa
26 ive study of three existing methods used for dead-time correction and background subtraction in neutr
27 ommended injected activity/body weight, peak dead-time correction factor, counting rates, and residua
31 less than 10% bias, from which corresponding dead-time, counting rates, and/or injected activity limi
32 ysis but, due to limitations in instrumental dead-times, discrimination of the "binding" and "base fl
38 development of CD signal in the stopped-flow dead time, indicative of the formation of a monomeric in
39 g speed by eliminating the image acquisition dead time induced by the beam flyback time combined with
40 azide, and phosphate accelerate decay of the dead time intermediate and for azide or fluoride lead di
41 ation occurs in the rate-determining step of dead time intermediate decay and that neither of the con
43 rous active site with superoxide generates a dead time intermediate whose absorption spectrum is iden
44 side-chain of F48W has lower mobility in the dead-time intermediate state than in both the fully dena
48 ed MBq/kg values are respected to limit peak dead-time losses during the bolus first-pass transit.
49 Using deuterated water as the unretained dead time marker for water-rich eluents combined with th
53 iple images at present is 1 ms/image, with a dead time of 3.2 ms between images, which will limit the
58 h colorimetric reactions showed the combined dead time of mixing and freeze-quenching to be submillis
59 shifted to 600 nm upon NAD(+) binding in the dead time of mixing of the stopped-flow instrument and r
60 lute head pressure of 85 psi, resulting in a dead time of only t(o) = 26 ms ( approximately 1900 cm/s
63 ocesses that are complete within the 5-10 ms dead time of stopped flow experiments account for the ma
64 g of the monomer were complete in the mixing dead time of stopped-flow CD and fluorescence spectrosco
67 ncrease in fluorescence within the 70-micros dead time of the continuous-flow experiment is consisten
69 Heme b and CuB were reduced within the 10-ms dead time of the freeze-quench experiment and remained a
70 olamine as the substrate occurred within the dead time of the instrument whenever coenzyme B(12) was
71 ermediate, with lambda(max) = 490 nm, in the dead time of the instrument, which then decays, with k =
72 hin 15 micros of its initiation and that the dead time of the measurement is 45 +/- 5 micros, which r
73 tate of cyt c, which is populated within the dead time of the mixer (<10 mus) and has a characteristi
74 mplete mixing was achieved within the mixing dead time of the mixer (20 micros), and the first observ
75 nism with a rapid phase occurring within the dead time of the spectrometer (<0.5 ms) followed by a si
76 solved fluorescence change during the 1.5 ms dead time of the stopped-flow experiment (burst phase).
78 uced FAD and urocanate that forms within the dead time of the stopped-flow instrument (~1 ms), with f
79 iff base intermediate (species A) during the dead time of the stopped-flow instrument, followed by fo
80 ation of a 325 nm absorption peak within the dead time of the stopped-flow instrument, likely the ket
83 resolved fluorescence change during the 1-ms dead time of the stopped-flow refolding measurements, wh
84 le dimeric intermediate and dimerizes in the dead-time of a manual-mixing kinetic experiment ( approx
85 concentrations reduces the enzyme within the dead-time of a stopped-flow instrument at 5 degrees C, i
89 ociate to form a dimeric intermediate in the dead-time of the SF instrument (approximately 5 ms); thi
90 efficient for the species formed during the dead-time of the stopped-flow experiment than for the fu
91 lation half-reaction, but is observed in the dead-time of the stopped-flow in the L-alanine transamin
92 oxidized flavin absorbance formed within the dead-time of the stopped-flow instrument ( approximately
93 rge decrease in fluorescence during the 2-ms dead-time of the stopped-flow measurement (burst phase)
97 pidFLIM was introduced, exploiting ultra-low dead-time photodetectors together with rapid electronics
99 mum tubings and show the complex relation of dead time, retention time, efficiency, and optimum veloc
100 ultrafast continuous-flow mixer (150 micros dead time) reveal that heme attachment to the polypeptid
101 ed for radioactive decay during acquisition, dead time, source attenuation, and source geometry effec
104 adding (57)Fe plus a 5 to 15 min processing dead time, was a quadrupole doublet typical of nonheme h
107 e plating and liquid handling errors, reduce dead times within the analysis cycle, and allow for comp