コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 nolithic columns and carbon dioxide/methanol mobile phase.
2 C18 column with ACN-formic acid 0.1% as the mobile phase.
3 percritical fluid chromatography using a CO2 mobile phase.
4 nitrile: water (75:25, v/v) is proposed as a mobile phase.
5 red in-line ionic strength adjustment of the mobile phase.
6 tonitrile each containing formic acid as the mobile phase.
7 aqueous methanol was applied as the washing mobile phase.
8 phy (UTLC) plates in 1-2 min using a ternary mobile phase.
9 etonitrile/1% CH3COOH (82/18) as the loading mobile phase.
10 ample treatment other than dilution with the mobile phase.
11 ium acetate water solution and methanol as a mobile phase.
12 dium (-)-dibenzoyl-l-tartarate as the chiral mobile phase.
13 e column to rid the ion-pair reagents of the mobile phase.
14 nd phosphate buffered saline solution as the mobile phase.
15 ationary phase and bile salt solution as the mobile phase.
16 series using a gradient of hexane-2-propanol mobile phase.
17 sensor and a PDZ domain (MAGI-1 PDZ1) in the mobile phase.
18 5 min using NH(4)NO(3) solution at pH 8.8 as mobile phase.
19 ttributed to the presence of methanol in the mobile phase.
20 d chromatography using an anionic surfactant mobile phase.
21 nts directly into the analyte solution or LC mobile phase.
22 y changing the amount of acetonitrile in the mobile phase.
23 t originally present in the sample or in the mobile phase.
24 e LC sample injection and delivery of the LC mobile phase.
25 and concentration of organic modifier in the mobile phase.
26 rometry (LC-MS) with methanol as the organic mobile phase.
27 n of hydrophobic albumin in a highly aqueous mobile phase.
28 ry phase and methanol: water (98:02, v/v) as mobile phase.
29 data were acquired using acetonitrile as the mobile phase.
30 -bungarotoxin (alpha(x)beta(y) nAChR) to the mobile phase.
31 ups, and the ability to use a highly aqueous mobile phase.
32 onitrile in water with 0.1% formic acid as a mobile phase.
33 ation/immobilization (s/i) buffer and to the mobile phase.
34 eluted with a gradient acetate-acetonitrile mobile phase.
35 -polar compounds employing a totally organic mobile phase.
36 components using an aqueous, non-denaturing mobile phase.
37 h 2% acetic acid/methanol (96:4, v/v) as the mobile phase.
38 cetonitrile/0.1% trifluoroacetic acid as the mobile phase.
39 dryness and the residue was reconstituted in mobile phase.
40 le (ACN): 0.1% phosphoric acid (60:10:30) as mobile phase.
41 of mixed-solvent liquid chromatography (LC) mobile phases.
42 ating the separation temperature under basic mobile phases.
43 egrees C and zero magnesium concentration in mobile phases.
44 lectivity, using both organic and water-rich mobile phases.
45 en added (generally, 0.1%), were employed as mobile phases.
46 rly observed for peptides when using high-pH mobile phases.
47 retention of solutes with high water-content mobile phases.
48 aqueous ammonium formate and acetonitrile as mobile phases.
49 lammable solvents (e.g., hexane) are used as mobile phases.
50 e positive mode of ionization with common LC mobile phases.
51 on of liquefied CO(2) to conventional liquid mobile phases.
52 ary phases and step-gradient aqueous-ethanol mobile phases.
53 s to separate intact proteins using volatile mobile phases.
56 n chromatography (SEC) with an MS-compatible mobile phase (59% water, 40% isopropanol, 1% formic acid
59 anol, 14.5mM triethylamine, and 5% methanol (mobile phase A) and 385mM hexafluoro-2-propanol, 14.5mM
60 tion into plasma allowed use of organic-rich mobile phase, achieving species separation in 4 min.
63 rifluoroacetic acid (TFA) is a commonly used mobile phase additive in liquid chromatography-mass spec
68 ce of polar mobile-phase solutions or acidic mobile-phase additives (e.g., formic or trifluoroacetic
69 ILIC/MS method was optimized on the basis of mobile-phase additives and pH, followed by evaluation of
72 ss distribution of the compound between the "mobile phase" (air and dust particles settled on the car
73 t variations in analyte diffusivities in the mobile phase, analyte elution temperatures, optimal line
74 hyl acetate/formic acid (6:10:1, v/v) as the mobile phase and 1% vanillin hydrochloric solution as st
75 size exclusion chromatography (SEC) such as mobile phase and column interaction effects on protein s
76 arise due to the compressible nature of the mobile phase and detector idiosyncrasies to eliminate ba
78 M, determination rate, ca. 10 samplesh(-1), mobile phase and ISA solution consumption, ca. 2.6 mL pe
79 d 1.65 mL/min, with pure acetonitrile as the mobile phase and naphtho[2,3-a]pyrene as the retained co
85 has been made transferable between different mobile phases and instrument setups by using a suitable
86 the use of solvents other than those of the mobile phase, and (iii) the need to stop the mobile phas
88 the missing cofactor is added to the column mobile phase, and the enzyme converts substrate into pro
90 heir increasing of retention at organic-rich mobile phases (approximately >90% v/v for acetonitrile w
91 neither for the hydrodynamic pumping of the mobile phase as in high-performance liquid chromatograph
95 that nonspherical particles can both reduce mobile-phase band broadening and increase permeability c
97 particle: acetonitrile in the interparticle mobile phase, C(18)-chain associated acetonitrile, and a
98 aining peptide eluted out of LC in an acidic mobile phase can be rapidly reduced prior to MS analysis
99 on behavior of bases (cationic acids) in the mobile phases can be better predicted by using the pH(ab
100 articles settled on the carpet) and the "non-mobile phase" (carpet fibers and pad) and the removal ra
104 th peak distortions due to strongly retained mobile phase components in other modes of liquid chromat
108 amines were separated on C18 column using a mobile phase composed of ammonium formate buffer and ace
109 lysis employed silica gel-coated TLC plates, mobile phase composed of chloroform:methanol:water:25% a
110 the elution of compounds in a higher organic mobile phase composition (retention times were approxima
111 degrees C), flow rate (1.0-2.5 mL min(-1)), mobile phase composition and equilibration time (1-5 min
112 ssed as are the effects of mat thickness and mobile phase composition on the chromatographic properti
116 < linear < ccc, is constant upon changes of mobile phase composition, gradient slope, and plasmid si
118 ic conditions includes the column selection, mobile-phase composition, pH value, buffer type, and con
119 aration was done with isocratic elution of a mobile phase comprising water (with 0.5% formic acid) an
120 was accomplished with isocratic elution of a mobile phase comprising water and methanol (92:8 v/v) on
121 eversed phase HPLC-ICP-MS, such as pH of the mobile phase, concentration of ion pairing reagents, typ
124 n Hibar C18 column (250x4.6mm, 5micro) using mobile phase consist of acetonitrile: water (90:10, v/v)
126 tion on silica gel-coated HPTLC plates using mobile phase consisted of chloroform:methanol:acetone:25
128 -crystalline bilayer phase, and an extremely mobile phase consistent with small vesicles or micelles.
129 c separation was achieved on C8 column using mobile phase consisting of (A) methanol:acetonitrile (8:
130 mm, 5 mum) column by gradient elution with a mobile phase consisting of 0.1% trifluoroacetic acid (pH
131 ne in food supplements was performed using a mobile phase consisting of 100% methanol, column tempera
132 ned on a Chiralcel(R) OJ-RH column using the mobile phase consisting of 10mM aqueous ammonium acetate
133 hic separation under gradient elution with a mobile phase consisting of acetonitrile and trifluoroace
134 enyl column (50 x 2.0 mm) and supported by a mobile phase consisting of acetonitrile plus 0.1% formic
135 mm, 5 mum) column by gradient elution with a mobile phase consisting of ammonium acetate 0.05 M and a
137 achieved using a C18 column with a gradient mobile phase consisting of solvents A (0.1% formic acid
138 thylene bridged hybrid phenyl column using a mobile phase consisting of water and methanol containing
140 ntrary to previous applications in which the mobile phase contained water, the improvement in sensiti
142 The optimized CEC-ESI-MS conditions were mobile phase containing 90/10 ACN/2.5 mM Tris, pH 8, she
144 alyzed by reversed-phase HPLC method using a mobile phase containing acetonitrile:methanol:2-propanol
145 bolites was carried out isocratically with a mobile phase containing both positively and negatively c
147 ed-phase liquid chromatography (IPRP) with a mobile phase containing triethylamine (TEA) and hexafluo
148 lumn, acetonitrile-trifluoroacetic acid as a mobile phase, coupled with UV detector at 205 nm, was su
150 n the retention factor (K) and the pH of the mobile phase demonstrate that the binding of cis-diols t
151 the dependency of solute retention with the mobile-phase density, complicating linear extrapolation
152 ution with mixtures of methanol and water as mobile phases; detection wavelength was set at 240 nm fo
153 his metric appears to be a good indicator of mobile-phase dispersion in ordered packed bed media, inc
154 ction of the mass transfer resistance in the mobile phase due to a flatter flow profile and faster an
155 tered when moving from an aqueous to organic mobile phase during a gradient elution, a key analytical
156 mobile phase, and (iii) the need to stop the mobile phase flow at any time during the full analytical
157 meters such as column type, buffer solution, mobile phase flow rate and sample injection volume were
162 ns for the generation of Fe nanoparticles, a mobile phase for As adsorption currently not a part of r
163 An anionic micellar medium was added to the mobile phase for increasing the fluorescence intensity a
164 methanol/aqueous trifluoroacetic acid as the mobile phase for size exclusion chromatography-ESI-MS an
165 spectrometry detection sensitivity than TFA mobile phases for LC-MS-based characterization of biopha
168 iagrams as a function of column temperature, mobile phase gradient or a multifactorial combination in
169 imized for organic modifier, ionic modifier, mobile phase gradient, flow rate, column type, MS source
171 reagent, m-nitrobenzyl alcohol (m-NBA), into mobile phases greatly facilitates the analysis of acidic
172 purpose, the addition of thiosulfate to the mobile phase has been used to elute Ag(I) species from t
176 ition of these supercharging reagents to the mobile phase in liquid chromatography (LC)-MS/MS increas
177 e change in the refractive index (RI) of the mobile phase in very fast gradients causes extremely ser
178 imation of the volumes of the stationary and mobile phases in dynamic equilibrium with eluents of var
181 ither post-column or by incorporation in the mobile phase increases specificity for all of the MC whi
182 ion following an increase of acetonitrile in mobile phase) initially exhibited by perfluoroalkyl acid
184 ndent accumulation of pyrene from an aqueous mobile phase into the center of individual C18-chromatog
185 d from potential contamination); (IV) the LC mobile phase is of much lower viscosity with respect to
186 y obtained with ammonium tartrate in the HIC mobile phases is orthogonal to that of reverse phase chr
187 stability of RNA in the denaturing/basic IEX mobile phase, lay out a protocol to determine the on-the
189 dge RP-18e (Merck), 10x4.6mm, with a washing mobile phase (methanol:water, 92:8, (v/v)) at a flow rat
191 ts in SFC separations using a nontraditional mobile phase mixture consisting of ammonium hydroxide co
193 polarities amenable to pSFC with appropriate mobile-phase modifiers and additives under normal-phase
194 phenyl-2-oxoethanal as the negative-ion-mode mobile-phase modifiers for the analysis of peptides.
196 on a Synergi Hydro analytical column with a mobile phase of 0.02 M ammonium formate in water and ace
197 e quinolones were resolved in <22min using a mobile phase of 0.05M SDS - 7.5% 1-propanol - 0.5% triet
198 ones were eluted without interferences using mobile phase of 0.05M SDS/12.5% 1-propanol/0.5% triethyl
200 uorescence detection within 10min by using a mobile phase of acetonitrile and acetate buffer of pH 6.
201 ected to high-performance LC with an optimal mobile phase of acetonitrile-water containing 0.1% formi
203 used under the optimum conditions predicted: mobile phase of H2SO4 0.005 mol L(-1) solution, flow rat
206 th a C18 column (2.1 x 100 mm, 1.7 mum) with mobile phases of 0.1% formic acid in water and acetonitr
208 and the final methanol concentration in the mobile phase on the peak resolution and peak symmetry wa
211 ditions and subsequent stability analysis in mobile phase, PBS buffer, and rat serum of 12 aryl sulfo
212 mobile phase velocities, the friction of the mobile phase percolating through the column bed generate
215 reagent (IPR) concentration, counterion, and mobile phase pH on the quality of the RPIP-UPLC separati
217 c nature of proteins, the use of a different mobile phase pH was successful to provide altered select
218 on (<9.9%CV), recovery (83-99%), robustness (mobile phase pH, column temperature and flow rate) and s
219 hromatographic parameters (stationary phase, mobile phase pH, temperature, organic solvent, and gradi
222 rious modes of protein chromatography at any mobile-phase pH, which may have significant implications
225 r peak shapes, particularly when eluted with mobile phases preferred for electrospray ionization mass
226 er, the water/acetonitrile and water/acetone mobile phases produced the better chromatographic separa
229 he discovery of this glycine additive in TFA mobile phases provides a simple and conventional approac
232 ively high percentage of acetonitrile in the mobile phase, resulting in stable and high efficiency io
234 of amphiphilic (surfactant) molecules to the mobile-phase solution in order to bring about the retent
235 nary phase particles with acetonitrile-water mobile phase solutions by confocal Raman microscopy.
236 bilization, because of the presence of polar mobile-phase solutions or acidic mobile-phase additives
241 rabutylammonium for carboxylic acids) in the mobile phase suppresses the ESI-MS signals in the gas ph
242 uted using a gradient methanol/water/toluene mobile phase system at a flow rate of 0.5 mL min(-1).
246 ationary phases in conjunction with multiple mobile-phase systems, as applied to the separation of 45
247 ous attempt, the vaporization of solutes and mobile phase takes place at atmospheric pressure into a
249 e successive physical transformations of the mobile phase that take place in very high pressure liqui
250 in their ability to resolve glycoforms using mobile phases that are compatible with online liquid chr
251 ntial-oil "matrix", replacing it with the LC mobile phase (the GC system is more protected from poten
253 Indeed it was found that, at organic-rich mobile phases, the transfer from the mobile to the stati
254 olumn is eluted with a predominantly organic mobile phase, then solutes can be retained through hydro
256 cellar conditions using 1-2% (v/v) 1-butanol mobile phase to remove plasma proteins and concentrate t
257 r for the transfer of polar species from the mobile phase to residual silanols, and this reduced barr
258 nd to develop gel porosity in contact with a mobile phase ultimately affecting the chromatographic pe
261 e linear velocity and the composition of the mobile phase used in the second dimension, its initial o
263 xtures of small molecules are carried out at mobile phase velocities close to (for isocratic runs) or
264 ith very fine particles are operated at high mobile phase velocities, the friction of the mobile phas
266 ate height with an increase of around 15% in mobile phase velocity in nonretained measurements of Cou
267 al plates as a function of column length and mobile phase velocity is a surface (z direction) to the
269 d due to the influence of temperature on the mobile phase viscosity and on the equilibrium constant o
270 column, the temperature dependencies of the mobile phase viscosity, and the retention factor of the
271 urea, did provide an accurate measure of the mobile-phase volume but only over a limited range of elu
276 mns were compared and the composition of the mobile phase was optimized to achieve baseline separatio
278 obtained with a gradient method in which the mobile phase was water (0.1% acetic acid) as solvent A a
279 ality grades, postcolumn modification of the mobile phase) was investigated in a pragmatic and decisi
281 tion, where the chemical constituents of the mobile phase were modified stepwise during analysis, thi
285 Several liquid chromatography stationary and mobile phases were evaluated, and it was found that hydr
287 h of analytical column, and flow rate of the mobile phase, were optimised for five selenium species;
289 aration of organic solutes in purely aqueous mobile phases whereby retention is controlled through te
290 ies by anion exchange chromatography using a mobile phase which is a 1:10 dilution of the extracting
291 covered glycine as a simple additive for TFA mobile phases, which mitigates ion suppression through a
293 h MS detection (HIC-MS) utilizing an aqueous mobile phase with a low concentration of a volatile salt
294 general, androgens required a stronger UPLC mobile phase with a slower flow rate and ESI of the oppo
296 phobic interaction chromatography (HIC) uses mobile phases with high salt concentration that are not
297 rol was performed on silica gel 60 plates, 4 mobile phases with suitable separation properties and co
298 e-exclusion chromatography employing aqueous mobile phases with volatile salts at neutral pH combined
299 ime direct comparison of acidities of any LC mobile phases, with different organic additives, differe
300 adding a small amount (e.g., 2 mM) into TFA mobile phases without compromising the chromatographic p