コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 e were performed with a Goldmann applanation tonometer.
2 metry was performed with a modified Goldmann tonometer.
3 d DBA/2NNia mice was measured with a rebound tonometer.
4 almologist specialists to use the ICare HOME tonometer.
5 Goldmann applanation tonometry and a rebound tonometer.
6 influence IOP as measured with a non-contact tonometer.
7 ter (GAT) and supine position with a Perkins tonometer.
8 traocular pressure (IOP) was measured with a tonometer.
9 sing the Ocular Response Analyzer noncontact tonometer.
10 were killed, IOP was measured with a rebound tonometer.
11 ressure was measured using a TonoLab rebound tonometer.
12 ere examined for changes in IOP by a TonoLab tonometer.
13 ice were measured using a commercial rebound tonometer.
14 ads and measured twice weekly with a rebound tonometer.
15 was calculated with the help of an arterial tonometer.
16 d IOP was measured with Goldmann applanation tonometer.
17 IOP was monitored with an applanation tonometer.
18 rown Norway rats was measured with a rebound tonometer.
19 traocular pressure (IOP) was measured with a tonometer.
20 IOP was measured with a rebound tonometer.
21 intraocular pressure (IOP) using the rebound tonometer.
22 rebound tonometry and the Proview phosphene tonometer.
23 oelasticity, were measured with a noncontact tonometer.
24 fference in measurement of IOP between the 2 tonometers.
25 H were determined with saline-filled balloon tonometers.
26 dynamic and can be influenced by the use of tonometers.
27 d poorest agreement between PN and the other tonometers.
28 a small MD in the measured IOP between the 2 tonometers.
29 s of ocular biometrics, AHD, and IOP using 4 tonometers.
30 0.98) showed excellent agreement between the tonometers.
31 ontact lenses beyond current wearable ocular tonometers.
32 esult in significant differences among the 3 tonometers.
33 raobserver variability were observed for all tonometers.
34 agreement (95% limits) seemed to vary across tonometers: 0.2 mmHg (-3.8 to 4.3 mmHg) for the NCT to 2
37 (IOP) was measured with the Dynamic Contour tonometer and central corneal thickness (CCT) was also e
39 calculated from the deep learning-predicted tonometer and mire diameters using the Imbert-Fick formu
42 te the relationship between mean IOP by each tonometer and rates of visual field loss over time, whil
44 New tonometers such as the dynamic contour tonometer and the Proview tonometer do not appear to be
45 IOP and CCT were measured with a hand-held tonometer and ultrasound pachymetry on the first, third,
46 an analysis was used to compare each pair of tonometers and whether the difference between tonometers
47 P) was measured using a Goldmann applanation tonometer, and central corneal thickness (CCT) was measu
52 corneal conforming shape used on a Goldmann tonometer appears to be a more sensitive and reliable in
54 baseline, both with an applanation resonance tonometer (ART) and an ocular response analyzer (ORA), a
55 The IOP measurements were conducted with a tonometer at 2 to 4 hours after onset of the nocturnal a
59 applanating tips and reduced weights in the tonometer body from those used in humans and species wit
60 (CCT) and IOP values obtained with all three tonometers but only the IOP values detected with the ICa
64 rwell, MA), the Perkins handheld applanation tonometer (Clement Clarke, Harlow, UK), and the Ocular B
65 ncluding IOP measurement using a non-contact tonometer, corneal structure evaluation with Pentacam Sc
66 The IOP readings generated by the rebound tonometer correlated very well with the actual pressure
68 rmined using 86Rb+ as a tracer for K+; glass tonometers coupled to a gas mixing pump were used to equ
69 ore, single-use tonometer tips or disposable tonometer covers should be considered when treating pati
70 g a newly developed modified Goldmann convex tonometer (CT) 1 year after myopic laser refractive surg
74 he dynamic contour tonometer and the Proview tonometer do not appear to be influenced by central corn
75 ements obtained with the iCare IC200 rebound tonometer during natural sleep and under general anesthe
76 d comparisons, were included, representing 8 tonometers: dynamic contour tonometer, noncontact tonome
79 The availability of an easy-to-use, reliable tonometer for IOP measurements in mice will allow more e
82 upright position with a Goldmann Applanation Tonometer (GAT) and supine position with a Perkins tonom
83 as measured once with a Goldmann applanation tonometer (GAT) and twice by ART (ART1, ART2), in random
86 In this study, the gold standard Goldmann tonometer (GAT) was compared to ICP and Tono-Pen AVIA (T
87 lar pressure (IOP) with Goldmann applanation tonometer (GAT) was measured at baseline and at 4 and 8
91 traocular pressure with Goldmann applanation tonometer (GAT-IOP), and endothelial cell density (ECD),
93 nse analyzer, Ocuton S, handheld applanation tonometer (HAT), rebound tonometer, transpalpebral tonom
94 acceptability and feasibility of iCare HOME tonometer (HT) and Virtual Field (VF) devices in the hom
95 vailable (45%), and otherwise with the iCare tonometer (iCare USA, Raleigh, NC), the Tono-Pen (Reiche
99 udy evaluated the applicability of a rebound tonometer in measuring intraocular pressure (IOP) in rat
100 Intraocular pressure (IOP) was measured by a tonometer in rats under ketamine-xylazine anesthesia.
101 g mechanical ventilation, with a nasogastric tonometer in situ, in whom enteral feeding was initiated
103 erformed using a modified electronic Schiotz tonometer in two positions: seated position, 70 degrees
105 eyes were monitored by slit lamp, a handheld tonometer, indirect ophthalmoscopy, electroretinography
108 uation, when measuring IOP with the Goldmann tonometer, it is likely that IOP is underestimated in ey
109 tonometers, one should consider that not all tonometer measurements are interchangeable and the propa
110 erwent Icare HOME (Icare Oy, Vanda, Finland) tonometer measurements to record IOP 4 times daily for 5
113 eters: dynamic contour tonometer, noncontact tonometer (NCT), ocular response analyzer, Ocuton S, han
114 [SD] 11.3+/-3.0 years) using three different tonometers: non-contact (NCT), the ICare and Goldmann ap
115 , representing 8 tonometers: dynamic contour tonometer, noncontact tonometer (NCT), ocular response a
117 ke a comparative analysis of three different tonometers on a group of healthy children to see whether
119 fore, when comparing IOP data from different tonometers, one should consider that not all tonometer m
120 thickness was assessed using an applanation tonometer or high-resolution MRI (23.4 microm2 in-plane)
122 anipulated by changing reservoir height, and tonometer pressure readings were recorded by an independ
123 measured by a standard Goldmann applanation tonometer prism (IOPg) and a modified correcting applana
126 of various disinfection methods for reusable tonometer prisms in eye care and to highlight how disinf
128 ) was measured on each eye using a new iCare tonometer probe, an iCare probe previously used and disi
134 s had their IOPs measured with iCare rebound tonometer (RT), ocular response analyzer corneal-compens
135 measured hourly using a portable noncontact tonometer (session with nocturnal hourly awakening).
136 only the IOP values detected with the ICare tonometer showed a statistically significant correlation
139 (IOPg) and a modified correcting applanation tonometer surface Goldmann prism (IOPc) before and after
140 hysician, immediately followed by ICare HOME tonometer (TA022, ICare Oy) measurement by the patient.
143 these instruments: the hand-held applanation tonometer, the Tono-pen XL, and the pneumatonometer.
151 the first experience of the use of a digital tonometer to understand the biomechanical properties of
152 Venous blood from healthy volunteers was tonometered to create different oxygen tensions simulati
153 d on many cohorts of children with different tonometers to determine how the values correlate between
156 ability and agreement between transpalpebral tonometers (TTs) and Goldmann applanation tonometer (GAT
158 he Tono-pen XL and the hand-held applanation tonometer underestimated the IOP, whereas the pneumatono
162 Repeatability of the measurements with each tonometer was assessed by calculating the mean and varia
170 agreement in IOP measurements from the four tonometers was analyzed with Data analysis using Bootstr
175 adjusted for proparacaine's effect on IOP, 3 tonometers were equivalent between each other when deter
185 three times using a Nidek NT-510 non-contact tonometer within a maximum of 5 min before and after VF