コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 the tryptophan fluorescence lifetime (60 mus dead time).
2 ted association reaction in the stopped-flow dead time.
3 higher column outlet pressures with minimal dead time.
4 values in a "burst phase" within the mixing dead time.
7 ssociated with refolding occurred during the dead time (4 ms) of the stopped-flow instrument, suggest
8 ly verify a theory that takes the effects of dead time, afterpulsing, and the finite sampling time on
10 high-speed separations by reducing detector dead time and by shifting optimal carrier gas velocity t
11 ith an assay that minimized the experimental dead time and which allowed for detection of N-acetyltyr
13 that takes nonideal detector effects such as dead-time and afterpulsing into account is developed and
14 information up to the start of the saccadic dead time, and (3) variability in saccade latency does n
15 d improvements in pulse shaping, integration dead time, and triggering, has an improved count-rate ca
20 d that the population formed during the 4 ms dead time contained multiple species that are stabilized
21 ommended injected activity/body weight, peak dead-time correction factor, counting rates, and residua
25 less than 10% bias, from which corresponding dead-time, counting rates, and/or injected activity limi
26 ysis but, due to limitations in instrumental dead-times, discrimination of the "binding" and "base fl
31 development of CD signal in the stopped-flow dead time, indicative of the formation of a monomeric in
32 azide, and phosphate accelerate decay of the dead time intermediate and for azide or fluoride lead di
33 ation occurs in the rate-determining step of dead time intermediate decay and that neither of the con
35 rous active site with superoxide generates a dead time intermediate whose absorption spectrum is iden
36 side-chain of F48W has lower mobility in the dead-time intermediate state than in both the fully dena
40 ed MBq/kg values are respected to limit peak dead-time losses during the bolus first-pass transit.
41 Using deuterated water as the unretained dead time marker for water-rich eluents combined with th
45 iple images at present is 1 ms/image, with a dead time of 3.2 ms between images, which will limit the
49 h colorimetric reactions showed the combined dead time of mixing and freeze-quenching to be submillis
50 shifted to 600 nm upon NAD(+) binding in the dead time of mixing of the stopped-flow instrument and r
51 lute head pressure of 85 psi, resulting in a dead time of only t(o) = 26 ms ( approximately 1900 cm/s
54 ocesses that are complete within the 5-10 ms dead time of stopped flow experiments account for the ma
55 g of the monomer were complete in the mixing dead time of stopped-flow CD and fluorescence spectrosco
56 ncrease in fluorescence within the 70-micros dead time of the continuous-flow experiment is consisten
58 Heme b and CuB were reduced within the 10-ms dead time of the freeze-quench experiment and remained a
59 olamine as the substrate occurred within the dead time of the instrument whenever coenzyme B(12) was
60 ermediate, with lambda(max) = 490 nm, in the dead time of the instrument, which then decays, with k =
61 hin 15 micros of its initiation and that the dead time of the measurement is 45 +/- 5 micros, which r
62 tate of cyt c, which is populated within the dead time of the mixer (<10 mus) and has a characteristi
63 mplete mixing was achieved within the mixing dead time of the mixer (20 micros), and the first observ
64 nism with a rapid phase occurring within the dead time of the spectrometer (<0.5 ms) followed by a si
65 solved fluorescence change during the 1.5 ms dead time of the stopped-flow experiment (burst phase).
67 iff base intermediate (species A) during the dead time of the stopped-flow instrument, followed by fo
68 ation of a 325 nm absorption peak within the dead time of the stopped-flow instrument, likely the ket
71 resolved fluorescence change during the 1-ms dead time of the stopped-flow refolding measurements, wh
72 le dimeric intermediate and dimerizes in the dead-time of a manual-mixing kinetic experiment ( approx
73 concentrations reduces the enzyme within the dead-time of a stopped-flow instrument at 5 degrees C, i
77 ociate to form a dimeric intermediate in the dead-time of the SF instrument (approximately 5 ms); thi
78 efficient for the species formed during the dead-time of the stopped-flow experiment than for the fu
79 lation half-reaction, but is observed in the dead-time of the stopped-flow in the L-alanine transamin
80 oxidized flavin absorbance formed within the dead-time of the stopped-flow instrument ( approximately
81 rge decrease in fluorescence during the 2-ms dead-time of the stopped-flow measurement (burst phase)
86 mum tubings and show the complex relation of dead time, retention time, efficiency, and optimum veloc
87 ultrafast continuous-flow mixer (150 micros dead time) reveal that heme attachment to the polypeptid
92 e plating and liquid handling errors, reduce dead times within the analysis cycle, and allow for comp
WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。