コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 ontaminated pharyngeal secretions around the endotracheal tube.
2 a face mask that prevented rebreathing or by endotracheal tube.
3 ry to generate an audible airleak around the endotracheal tube.
4 re made directly into the trachea through an endotracheal tube.
5 aced in the ventilation circuit, next to the endotracheal tube.
6 ots of P. multocida were administered via an endotracheal tube.
7 ered into the right atrium, femoral vein, or endotracheal tube.
8 ssure were measured at the distal end of the endotracheal tube.
9 asured with a pneumotachometer placed at the endotracheal tube.
10 bbits were tracheostomized with a multilumen endotracheal tube.
11 rogen dioxide was <1 ppm at the level of the endotracheal tube.
12 ving mechanical ventilation through a cuffed endotracheal tube.
13 nium in normal saline was instilled into the endotracheal tube.
14 of large amounts of sand from the patient's endotracheal tube.
15 red during ventilation with the conventional endotracheal tube.
16 by the use of a catheter passed through the endotracheal tube.
17 with the pressure at the carinal end of the endotracheal tube.
18 of the ventilator to the carinal end of the endotracheal tube.
19 to the test lung collected drug exiting the endotracheal tube.
20 nce imposed by the breathing circuit and the endotracheal tube.
21 ressure triggering at the carinal end of the endotracheal tube.
22 theter positioned 2 cm beyond the tip of the endotracheal tube.
23 days mostly (n = 432; 56%) ventilated via an endotracheal tube.
24 secretions from the interior surface of the endotracheal tube.
25 using a bronchial blocker or a double-lumen endotracheal tube.
26 stylet end positioned within the tip of the endotracheal tube.
27 as been used to confirm the placement of the endotracheal tube.
28 ed in ten dogs by partially constricting the endotracheal tube.
29 The model was applied to diagnose obstructed endotracheal tube.
30 significant pressure gradient exists across endotracheal tubes.
31 mall for commercially available double-lumen endotracheal tubes.
32 lling plastic devices, such as catheters and endotracheal tubing.
33 were initially intubated with a conventional endotracheal tube (2.5-mm internal diameter; 3.6-mm oute
34 d (1.98+/-1.68) in comparison with untreated endotracheal tubes (3.72+/-2.20, p=.045) or those treate
35 tubes used in patients compared with unused endotracheal tubes (5.4 +/- 0.7 vs. 6.0 +/- 0.6 mL, p <.
36 es used in patients compared with the unused endotracheal tubes (7.5 +/- 0.4 vs. 6.7 +/- 1.2 mm, p <.
37 iratory limb collected albuterol exiting the endotracheal tube and any albuterol lost, respectively.
38 record airway pressure just proximal to the endotracheal tube and cuff pressure via the pilot tube.
42 athing (work to spontaneously inhale through endotracheal tube and ventilator circuit), work by venti
47 s a clinical comparison between double-lumen endotracheal tubes and bronchial blockers to determine w
50 eal tube segment volumes, between the unused endotracheal tubes and the endotracheal tubes used in pa
51 to the lungs of anesthetized rats through an endotracheal tube, and a series of square-wave pulses we
52 ormedics 3100B high-frequency ventilator, an endotracheal tube, and a test lung, tidal volume was mea
53 del consisting of a HFOV and circuit, 4.5-mm endotracheal tube, and lung simulator was assembled.
54 pediatric breathing circuit, 4.0- or 6.0-mm endotracheal tube, and lung simulator, was assembled.
55 se of prophylactic probiotics and subglottic endotracheal tubes are cost-effective for preventing VAP
56 observed that women and patients with narrow endotracheal tubes are often successfully extubated with
57 ymorphism analysis of virus isolated from an endotracheal-tube aspirate and from bronchoalveolar lava
58 lmonary oxyradical stress were determined on endotracheal tube aspirates and were normalized accordin
60 tilation animals received perflubron via the endotracheal tube at either full functional residual cap
61 100B high-frequency ventilator and an 8.0-mm endotracheal tube attached to a 48.9-L plethysmograph.
62 Systemic treatment with linezolid limits endotracheal tube biofilm development and methicillin-re
64 thesia, 40 piglets (11.5+/-0.3 kg) underwent endotracheal tube clamping (6.8+/-0.3 minutes) until sim
68 The Mucus Shaver is helpful in preventing endotracheal tube colonization by potentially harmful mi
69 (PLV, n = 15) with perflubron (18 mL/kg via endotracheal tube), conventional mechanical ventilation
70 sure amplitude, frequency, and the use of an endotracheal tube cuff leak, and to maintain oxygenation
72 tidal volume, distal pressure transmission, endotracheal tube cuff leaks, and simulated clinical con
73 procedures relevant to HFOV include setting endotracheal tube cuff leaks, performing lung recruiting
75 nvestigator deflated and then reinflated the endotracheal tube cuff until tracheal seal was reestabli
79 a ventilator "ON" at the carinal end of the endotracheal tube decreases imposed work of breathing by
81 ude transmission is directly correlated with endotracheal tube diameter and peripheral airway resista
82 r lobe atelectasis believed to be related to endotracheal tube displacement during the passage of the
83 gery, specifically the use of a double-lumen endotracheal tube (DLT); a few centers use carbon dioxid
84 initiating mechanical ventilation through an endotracheal tube during a 12-wk interval formed the stu
86 The volume of secretions collected on the endotracheal tube during the 1.5-hr experiment and on a
87 anically ventilated for 4 hrs, with head and endotracheal tube elevated 30 degrees from horizontal.
88 (combined nasal wash [NW]-throat swab [TS], endotracheal tube [ET] aspirate, or bronchoalveolar lava
89 p were kept prone, intubated with a standard endotracheal tube (ETT), and mechanically ventilated for
90 is study was to determine the resistances of endotracheal tubes (ETTs) commonly used in neonatal and
96 k bacterial deposits were present on all the endotracheal tubes from the control group (p < .001 by F
97 n microscopy showed little secretions on the endotracheal tubes from the study group, whereas thick b
101 men, especially when breathing through small endotracheal tubes, have a higher f/VT (including likeli
104 Perfluorocarbon was then instilled via the endotracheal tube in an amount estimated to represent fu
105 as achieved by instilling perflubron via the endotracheal tube in an amount estimated to represent th
107 evaluated a new device designed to clean the endotracheal tube in mechanically ventilated patients, t
110 cal judgment and direct visualization of the endotracheal tube in the trachea are required to unequiv
111 ory disease; c) more critical positioning of endotracheal tubes in younger patients and their movemen
113 lung compliance (10, 30, and 50 mL/cm H2O), endotracheal tube internal diameter (6, 7, and 8 mm), bi
118 ificant iatrogenic withdrawal, and unplanned endotracheal tube/invasive line removal were not signifi
119 alse-positive results mean that although the endotracheal tube is in the esophagus, the device indica
120 alse-negative results mean that although the endotracheal tube is in the trachea, the device indicate
124 anically ventilated patients with no audible endotracheal tube leak, measured endotracheal tube leak
126 tiple etiologies, including local effects of endotracheal tubes, neuromuscular weakness, and an alter
127 piratory flow demands and degrees of partial endotracheal tube occlusion (25%, 50%, and 75%) were stu
128 assed through the anaesthetized nose, and an endotracheal tube occlusion device was placed in the ven
134 unit who had a f/VT measured through an oral endotracheal tube (off of ventilatory support) during 1
135 ressure ventilation; the need to replace the endotracheal tube on admission to the PICU; and the use
136 n be achieved with the use of a double-lumen endotracheal tube or an independent bronchial blocker.
137 absolute lung separation with a double-lumen endotracheal tube or the use of a bronchial blocker in a
138 95% CI, 2.07-33.63] and interacting with the endotracheal tube or tracheotomy site [odds ratio, 5.15;
139 the endotracheal tube, it either entered the endotracheal tube or was lodged at the inflated endotrac
141 ation (n = 16) with perflubron (18 mL/kg via endotracheal tube) or conventional mechanical ventilatio
142 tion, lung secretion drainage, silver-coated endotracheal tubes) or oropharyngeal prophylactic method
143 vation, use of subglottic secretion drainage endotracheal tubes, oral care, chlorhexidine mouth care,
144 blish an airway is by placing a single-lumen endotracheal tube orally or nasotracheally while the pat
145 ly decreases the time to recognize misplaced endotracheal tubes placed during resuscitation, their us
149 unit and explored the measurement of twitch endotracheal tube pressure as a less invasive technique
151 ageal pressure was 0.93, and that for twitch endotracheal tube pressure to transdiaphragmatic pressur
156 ital perspective included subglottic suction endotracheal tubes, probiotics, and the Institute for He
157 modality can be utilized for confirmation of endotracheal tube.Recent interest in axillary vein cannu
161 The use of the ultrathin walled Kolobow endotracheal tube resulted in significant decreases in a
162 cantly (p < or = .05) greater for the 6.0-mm endotracheal tube, rigid spacer, dry air, and pulmonary
170 with administration of perflubron through an endotracheal tube sideport or conventional mechanical ve
171 bation with 1 of 2 high-volume, low-pressure endotracheal tubes, similar except for a silver coating
172 are coisolated from infections of catheters, endotracheal tubes, skin, eyes, and the respiratory trac
173 yethylene catheter was threaded through each endotracheal tube such that it could be positioned to me
174 thin 2 hrs of intubation to receive standard endotracheal tube suctioning treatment or standard sucti
176 5 cm H2O, spontaneous ventilation through an endotracheal tube (T piece), and pressure support ventil
177 were reintubated with a Univent single lumen endotracheal tube that incorporates an internal catheter
178 rom the trachea contiguous to the tip of the endotracheal tube, the distal trachea, the carina, and t
179 hen the model is used to diagnose obstructed endotracheal tube, the method has a sensitivity of 90% a
180 of bright red blood were suctioned from the endotracheal tube, there was no evidence of coagulopathy
182 The Mucus Shaver is advanced to the distal endotracheal tube tip, inflated, and subsequently withdr
184 /mmol) was administered to normal rabbits by endotracheal tube to assess biodistribution, route of el
186 and the lowest bacterial burden was found in endotracheal tubes treated with linezolid (1.98+/-1.68)
187 egments was also statistically smaller among endotracheal tubes used in patients compared with the un
188 ent volumes were statistically smaller among endotracheal tubes used in patients compared with unused
189 etween the unused endotracheal tubes and the endotracheal tubes used in patients, was 9.8% (range, 0-
195 ) were measured (period 1), the conventional endotracheal tube was replaced with a Kolobow tube, and
197 at an amplitude of 80 cm H2O) when a 3.5-mm endotracheal tube was used; however, gas flow was attenu
198 during the experiment and the length of the endotracheal tubes was the same for both the conventiona
200 ted patients and only 8% of patients with an endotracheal tube were mobilized out of bed as part of r
202 colonized, whereas in the control group ten endotracheal tubes were colonized (8% vs. 83%; p < .001)
206 d and C(1) decreased with partial obstructed endotracheal tube, whereas R(1) increased and L and C(2)
207 ternative technique is to use a double-lumen endotracheal tube while using an airway catheter exchang
210 ll were mechanically ventilated via a cuffed endotracheal tube, with ventilator rate and tidal volume
WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。