戻る
「早戻しボタン」を押すと検索画面に戻ります。

今後説明を表示しない

[OK]

コーパス検索結果 (1語後でソート)

通し番号をクリックするとPubMedの該当ページを表示します
1 ocks for sinus tachycardia or noise/artifact/oversensing.
2 gh impedance and/or nonphysiological "noise" oversensing.
3  triggered either by high impedance or rapid oversensing.
4 ly present as inappropriate shocks caused by oversensing.
5 f right atrial sensing (1%), and ventricular oversensing (0.2%).
6 ), sinus or atrial tachycardias (9%), and/or oversensing (4%).
7 ested an algorithm that uses two measures of oversensing and one measure of abnormal impedance to det
8  electrograms (EGMs) alerts the physician to oversensing and undersensing problems, which may manifes
9 ation rules, enhancements to minimize T-wave oversensing, and features that restrict therapy to regul
10 ors to reduce inappropriate shocks for rapid oversensing caused by conductor fractures and reported f
11                                              Oversensing combined with abnormal impedance trends may
12 rdia (HR: 0.97, p = 0.86) and noise/artifact/oversensing (HR: 0.91, p = 0.76) was comparable to that
13 escribed mode of ICD lead failure: prolonged oversensing immediately after shock therapy.
14 sures were met or abnormal impedance and one oversensing measure occurred.
15                                          The oversensing measures consisted of a counter for RR inter
16                                      The two oversensing measures used in the algorithm predicted 72%
17      Lead failures were identified when both oversensing measures were met or abnormal impedance and
18 ling at least two of the three impedance and oversensing measures, the sensitivity of our algorithm w
19                                              Oversensing most often manifested as inhibition, althoug
20 e of lead failure for the 6936 consisting of oversensing of electrical noise following shocks, 3) ear
21 diation at typical clinical doses results in oversensing of ICRMDs in the majority of devices tested,
22 used by supraventricular tachyarrhythmias or oversensing of signals.
23 anted defibrillators tested were affected by oversensing of the electric field as verified by telemet
24  ICDs tested in 41 patients were affected by oversensing of the EMI field of the cellular telephones
25 nt and/or physician when triggered by either oversensing or excessive increases in impedance.
26 g only ICD diagnostics identifies leads with oversensing or high impedance as fractures or connection
27                                              Oversensing was observed in 20 of 21 devices at maximum
28                                              Oversensing was transient and ceased as soon as the devi
29 d inappropriate shocks, mainly due to T-wave oversensing, which was mostly solved by a software upgra
30 er extremely high maximum impedance or noise oversensing with a normal impedance trend indicated a fr

WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。