コーパス検索結果 (1語後でソート)
通し番号をクリックするとPubMedの該当ページを表示します
1 s of bipolar electrode spacing on PNS and LV pacing thresholds.
2 se the PNS threshold without compromising LV pacing thresholds.
3 MRI) sensing amplitudes, lead impedances, or pacing thresholds.
4 (COI) that may result in acute elevation of pacing thresholds.
5 venous system while providing low and stable pacing thresholds.
6 s not significantly affect defibrillation or pacing thresholds.
7 lator function by raising defibrillation and pacing thresholds.
9 gmentation of the electrogram and increasing pacing threshold (3.1 +/- 3.3 mV to 7.1 +/- 3.8 mV, p <
11 Regular measurements of lead impedance and pacing thresholds allow for early detection of lead insu
14 not always possible because of the elevated pacing threshold and the low voltage and fractionated po
17 ay explain changes in the defibrillation and pacing thresholds and the difficulty of lead extraction
18 8 mV, respectively; median acute and chronic pacing thresholds at 0.5 ms were 0.9 and 0.7 V, respecti
19 t a COI, 5 dislodged acutely and 15 had high pacing thresholds at 10 min, requiring repositioning.
21 llow-up were, respectively, as follows: mean pacing threshold (at a 0.4-ms pulse width), 0.40 +/- 0.2
22 ntricular steroid leads had improved chronic pacing thresholds (at 2 years: for steroid leads, 1.9 mu
25 layed effects including signal interference, pacing threshold changes, and premature battery depletio
26 ode spacing was reduced (P<0.05), whereas LV pacing thresholds did not change significantly (P>0.05).
27 pacing performance (sensing, impedance, and pacing threshold) either improved or were stable within
28 in 1.7%), cardiac perforation (in 1.3%), and pacing-threshold elevation requiring percutaneous retrie
30 eine (200 mumol/L) significantly reduced the pacing threshold for both SR Ca(2+) and APD alternans (1
31 Ca(2+) alternans, but not APD alternans, the pacing threshold for discordance, or threshold for VF.
32 (p < 0.001), with subsequent improvement in pacing thresholds from 1.5 +/- 0.4 V to 0.8 +/- 0.3 V (p
35 tal fluoroscopy, lateral roentgenograms, and pacing threshold levels were studied in two groups of pa
36 ry efficacy end point was both an acceptable pacing threshold (</=2.0 V at 0.4 msec) and an acceptabl
38 though electrogram amplitude correlated with pacing threshold (r=0.64, P<0.0001), many isthmuses had
43 C antiarrhythmic drug, on defibrillation and pacing thresholds were evaluated in patients undergoing
45 OI indicates that within 10 min of fixation, pacing threshold will return to an acceptable range even
WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。