戻る
「早戻しボタン」を押すと検索画面に戻ります。

今後説明を表示しない

[OK]

コーパス検索結果 (1語後でソート)

通し番号をクリックするとPubMedの該当ページを表示します
1  energy toward triplet states, enhancing the phosphorescence.
2 o produce surprisingly efficient solid-state phosphorescence.
3 ient and reversible quenching of the (3)MLCT phosphorescence.
4 to downward, resulting in the suppression of phosphorescence.
5 e cortex (pCO2) was measured by quenching of phosphorescence.
6 croM), whereas Ca2+ showed no effect on Tb3+ phosphorescence.
7 decreases in flexibility, as measured by Trp phosphorescence.
8 esterases were investigated using tryptophan phosphorescence.
9 1,3]thiadiazoles and why they are capable of phosphorescence.
10 DF), whereas compound 2 shows a pure, yellow phosphorescence.
11  states can be monitored by room-temperature phosphorescence.
12 omplex formation produces a red-shift of the phosphorescence 0, 0-band (DeltaE0,0) of Trp37 as well a
13  nucleic acids results in a red-shift of the phosphorescence 0,0-band (delta E(0,0)) of the aromatic
14 ily access and emit from its T1 state with a phosphorescence (3)(7a)* lifetime of tauP = 395 mus at 7
15 ganometallic systems, where ligand-localized phosphorescence ((3) pi-pi*) is mediated by ligand-to-me
16                                   Long-lived phosphorescence (4 s at 77 K) was recorded, and quantum-
17  can subsequently be detected by its 1270 nm phosphorescence (a(1)Delta(g) --> X(3)Sigma(g)(-)) with
18 sistent with expectations from the theory of phosphorescence, an inverse correlation between out-of-p
19                                              Phosphorescence analyzer that measured dissolved O2 as f
20 culations corroborate that the emissions are phosphorescence and arise from charge transfer (LML'CT)
21 stable palladium complexes that exhibit both phosphorescence and delayed fluorescence are developed.
22 g red metal-to-ligand-charge-transfer (MLCT) phosphorescence and electrophosphorescence.
23 SC rate modulates the intensity ratio of the phosphorescence and fluorescence emission bands, with po
24  These results are in general agreement with phosphorescence and ODMR measurements monitoring W.
25                                              Phosphorescence and optical detection of magnetic resona
26 oly(I), and rG(8)] have been investigated by phosphorescence and optically detected magnetic resonanc
27                                              Phosphorescence and optically detected magnetic resonanc
28                                              Phosphorescence and optically detected magnetic resonanc
29 and polynucleotides has been investigated by phosphorescence and optically detected magnetic resonanc
30 ose agreement between the quenching of donor phosphorescence and the efficiency of resonance energy t
31                                          The phosphorescence and zero field optically detected magnet
32 effects, experiments utilizing endoperoxide, phosphorescence, and chemiluminescence quenching studies
33  Photophysical studies such as fluorescence, phosphorescence, and laser flash photolysis in addition
34                                Fluorescence, phosphorescence, and optical detection of triplet state
35 element has been studied using fluorescence, phosphorescence, and optically detected magnetic resonan
36  states was estimated from the time-resolved phosphorescence anisotropy (TPA) decay of SR in which Ca
37                   We have used time-resolved phosphorescence anisotropy (TPA) of actin to evaluate do
38                                Time-resolved phosphorescence anisotropy (TPA) of ErIA and steady-stat
39 otion of actin was measured by time-resolved phosphorescence anisotropy (TPA) of erythrosin iodoaceta
40                       We have used transient phosphorescence anisotropy (TPA) to detect changes in ac
41                   We have used time-resolved phosphorescence anisotropy (TPA) to probe rotational dyn
42                   We have used time-resolved phosphorescence anisotropy (TPA) to study the rotational
43                                Time-resolved phosphorescence anisotropy (TPA) was used to determine c
44 dynamics of actin, detected by time-resolved phosphorescence anisotropy (TPA).
45 d to C374 on actin, as detected by transient phosphorescence anisotropy (TPA).
46 in-iodoacetemide (ErIA), using time-resolved phosphorescence anisotropy (TPA).
47                                Time-resolved phosphorescence anisotropy and fluorescence resonance en
48 tional mobility of IL-1 RI was assessed with phosphorescence anisotropy decay measurements using eryt
49 ecrease in the correlation time of transient phosphorescence anisotropy decays.
50                                Time-resolved phosphorescence anisotropy experiments demonstrated that
51             Complementary measurements using phosphorescence anisotropy of erythrosin isothiocyanate
52    A strong positive correlation between the phosphorescence anisotropy of F-actin under specific con
53                             The steady-state phosphorescence anisotropy of skeletal tropomyosin on F-
54                            We used transient phosphorescence anisotropy to detect the microsecond rot
55                   We have used time-resolved phosphorescence anisotropy to investigate the effects of
56                       We have used transient phosphorescence anisotropy to monitor the microsecond ro
57                    Here, using time-resolved phosphorescence anisotropy, electron cryomicroscopy, and
58 We have used optical spectroscopy (transient phosphorescence anisotropy, TPA, and fluorescence resona
59 /- 9 degrees as determined from steady-state phosphorescence anisotropy.
60                             Fluorescence and phosphorescence are clearly discriminated using a picose
61 )dbm(I)PLA with weak fluorescence and strong phosphorescence are promising as 'turn on' sensors for a
62 e melt and provides evidence of the value of phosphorescence as a probe of dynamic site heterogeneity
63 ed by two methods: direct measurement of its phosphorescence at 1275 nm and chemical trapping using u
64 of fluorescence and appearance of structured phosphorescence at 77 K are attributed to nitrophenyl-lo
65 target molecule, allowing the observation of phosphorescence at room temperature and alleviating cons
66                                          The phosphorescence band is blue-shifted ca. 20 nm in the ag
67  RT) and coarse- and fine-tuning to multiple phosphorescence bands across the visible spectrum via lu
68  application by integrating the sensors of a phosphorescence based CGM system into a standard insulin
69                                              Phosphorescence-based measurements can be used to determ
70 eral reaction scheme to the development of a phosphorescence-based sensing system for cyanogen halide
71       In this study, the authors developed a phosphorescence-based technique for measuring corneal O2
72      Heavy-atom substitution alone increases phosphorescence by a given, not variable amount.
73  energy transfer, minimized quenching of the phosphorescence by electron transfer and increased signa
74       Here we report bright room temperature phosphorescence by embedding a purely organic phosphor i
75 e the theory and principles of computational phosphorescence by highlighting studies of classical exa
76  it is shown that both chemiluminescence and phosphorescence can also be observed in a highly directi
77  studies reveal that bright room temperature phosphorescence can be realized in purely organic crysta
78 r (IPr --> AuM2) and interligand (IPr --> E) phosphorescence character, as revealed by time-dependent
79                                              Phosphorescence data also suggest that the N- and C-term
80                                              Phosphorescence data are consistent with heavy-atom assi
81 ed experiments, monitoring the 1270 nm (1)O2 phosphorescence decay generated upon laser irradiation a
82                                 The rates of phosphorescence decay of 4,7-dimethylindanone (2), 6,9-d
83 ge ODMR and of short-lived components in the phosphorescence decay of the complex of R6WGR6 with poly
84  determined by observing their effect on the phosphorescence decay of the triplet state of rose benga
85 s a substrate, with [O(2)] obtained from the phosphorescence decay rate of a palladium phosphor.
86 dentical lifetimes to those observed for the phosphorescence decays when measured under identical exp
87           This discovery opens up the use of phosphorescence-detected hydrogen exchange as a sensitiv
88 n designed to couple the aggregation induced phosphorescence, displayed by the core in the solid stat
89                        It was found that the phosphorescence efficiency depends on the orientation of
90 onless transitions and hence greatly enhance phosphorescence efficiency of metal-free organic materia
91 erature triplet emitters are correlated with phosphorescence efficiency.
92 dence for this comes from a fast rise in the phosphorescence emission and measurements of a correspon
93  was monitored by measuring the steady-state phosphorescence emission anisotropy (rFA) of the triplet
94  filaments were monitored using steady-state phosphorescence emission anisotropy of the triplet probe
95                                  Data on the phosphorescence emission energy and lifetime from erythr
96    Single molecules are detected through the phosphorescence emission of their triplet states.
97 studies together with fluorescence and (1)O2 phosphorescence emission quantum yields collected on Br2
98                                              Phosphorescence emission spectra are resolved and shift
99  radiative decay, which in turn boosts (1)O2 phosphorescence emission to a greater extent.
100                   Bathochromic shifts in the phosphorescence emission upon exciting to the red in CT
101 was projected at an angle on the retina, and phosphorescence emission was imaged after intravitreal i
102 onal temperature sensing agents, (ii) bright phosphorescence emission, (iii) a reversible thermal res
103  anthraniloyl and further enhancement in the phosphorescence emitted by Tb3+ upon excitation at 320 n
104  energy transfer, solid-state solvation, and phosphorescence enables 10-fold increases in the power o
105 al excimer geometry and the magnitude of the phosphorescence energy lowering in going from the monome
106 erligand distances around 3.5-3.8 A, lead to phosphorescence energy lowerings with respect to the mon
107                          We describe the red phosphorescence exhibited by a class of structurally sim
108   The absorption, steady-state fluorescence, phosphorescence, fluorescence lifetime, and phosphoresce
109 eport a strategy for modulating fluorescence/phosphorescence for a single-component, dual-emissive, i
110 locene (Cp)(2)Ti(NCS)(2) exhibits an intense phosphorescence from a ligand-to-metal charge transfer t
111 ons were monitored by measuring the decay of phosphorescence from a Pd phosphor in solution; the deca
112 de polymer is water soluble, and it exhibits phosphorescence from a triplet pi,pi exciton based on th
113                  This study illustrates that phosphorescence from erythrosin B is sensitive both to l
114                                          The phosphorescence from Pt-p is quenched by viologens with
115 lar interactions to enhance room-temperature phosphorescence from purely organic materials in amorpho
116                                              Phosphorescence from the excited PS was quenched by the
117                                    Polarized phosphorescence from the triplet probe erythrosin-5-iodo
118 for the direct signature of singlet fission, phosphorescence from the triplet state, in a model polym
119 ade use of direct time-resolved detection of phosphorescence, having the ability to efficiently rejec
120       A previously developed optical section phosphorescence imaging system was used to measure P(O2)
121                                     However, phosphorescence in organic molecules is rare at room tem
122  display highly efficient blue or blue-green phosphorescence in solution (Phi = 0.41-0.87) and the so
123 ment complexes, and room-temperature near-IR phosphorescence in the case of several 5d metal complexe
124 i-diboryne compounds (n = 2, 3) show intense phosphorescence in the red to near-IR region from their
125 nds responsible for pre-edge fluorescence or phosphorescence in the visible.
126                         The strongly allowed phosphorescence in these complexes is the result of sign
127 be challenging to vary relative fluorescence/phosphorescence intensities for practical sensing applic
128 eight polymer with balanced fluorescence and phosphorescence intensities serve as ratiometric tumour
129                                          The phosphorescence intensity (lifetime), emission energy, a
130                                          The phosphorescence intensity decay of PdOEP in the polymer
131  The cofilin concentration-dependence of the phosphorescence intensity is sigmoidal, consistent with
132                         Cofilin quenches the phosphorescence intensity of actin-bound ErIA, indicatin
133  resolution, using the temperature-dependent phosphorescence intensity of the rare earth chelate Eu-T
134 rate was obtained by fitting the tail of the phosphorescence intensity profile to an exponential.
135 onstruct exhibits concomitant changes in its phosphorescence intensity ratio and phosphorescence life
136              In contrast to the reduction in phosphorescence intensity, the changes in the rates of r
137                                              Phosphorescence is a phenomenon of delayed luminescence
138                Oxygen-dependent quenching of phosphorescence is a useful and essentially noninvasive
139                                              Phosphorescence is among the many functional features th
140      The bridging of the spin prohibition in phosphorescence is commonly analyzed by perturbation the
141 k deoxygenation system for measuring protein phosphorescence is described.
142                                  Because the phosphorescence is effectively quenched by molecular oxy
143                              Ir(III) corrole phosphorescence is observed at ambient temperature at wa
144                                              Phosphorescence is observed from these clusters in glass
145                                              Phosphorescence is the simplest physical process which p
146                          The single molecule phosphorescence is very sensitive to molecular oxygen.
147 loss of triplets, a key process to achieving phosphorescence, is difficult without heavy metal atoms.
148 riplets and exhibits efficient near-infrared phosphorescence (lambda(em) = 772 nm, Phi = 0.26).
149                      Careful analysis of the phosphorescence lifetime distribution of Trp 109 in Esch
150          Retinal PO2 maps were computed from phosphorescence lifetime images, and oxygen profiles thr
151 ension was measured in retinal vessels using phosphorescence lifetime imaging and converted to arteri
152  a melanoma tumour spheroid using one-photon phosphorescence lifetime imaging microscopy (PLIM) and a
153                           An optical section phosphorescence lifetime imaging system was developed fo
154 ing our previously developed optical section phosphorescence lifetime imaging system.
155 n innovative optical system for dual oxyphor phosphorescence lifetime imaging to near-simultaneously
156                                              Phosphorescence lifetime imaging was used to measure eac
157                                              Phosphorescence lifetime imaging was used to measure flu
158 ot only simple lifetime measurement but also phosphorescence lifetime imaging.
159 ield optical microscopy, including 2D and 3D phosphorescence lifetime imaging.
160 s in its phosphorescence intensity ratio and phosphorescence lifetime in response to copper(II) ion.
161        The room-temperature tryptophan (Trp) phosphorescence lifetime is sensitive to details of the
162  phosphorescence, fluorescence lifetime, and phosphorescence lifetime measurements were carried out.
163     A frequency-domain approach was used for phosphorescence lifetime measurements.
164 s of metalloporphyrins, including two-photon phosphorescence lifetime microscopy (2PLM) and two-photo
165             Recent development of two-photon phosphorescence lifetime microscopy (2PLM) of oxygen ena
166                             Using two-photon phosphorescence lifetime microscopy, we determined the a
167 mercial time-resolved fluorescence reader in phosphorescence lifetime mode.
168 isoforms also differ in their effects on the phosphorescence lifetime of the actin-bound erythrosin i
169 to O2 (RSD at 21 KPa 1.9%), and reproducible phosphorescence lifetime readings.
170 luctuations on the picosecond time scale and phosphorescence lifetime was observed.
171  excitation and systematic variations in the phosphorescence lifetime with wavelength indicated that
172 ameters D and E, and a significantly reduced phosphorescence lifetime, each consistent with aromatic
173                                          The phosphorescence lifetime-based measurement circumvents t
174 scular P(O2) was measured by determining the phosphorescence lifetime.
175 et lifetimes were confirmed by measuring the phosphorescence lifetimes and with the help of diffusion
176 CT state increase from 4 to 12 ps, while the phosphorescence lifetimes are approximately 80 micros.
177                                          The phosphorescence lifetimes are shorter by an order of mag
178                                   Their long phosphorescence lifetimes in living cancer cells give ri
179                                              Phosphorescence lifetimes of 2 ms for N-acetyl-L-tryptop
180 the system was demonstrated by measuring the phosphorescence lifetimes of N-acetyl-L-tryptophanamide,
181 examined in detail, and compared with Trp102 phosphorescence lifetimes that were previously measured.
182 hotoluminescence and excitation spectra, and phosphorescence lifetimes, are presented.
183                              The homogeneous phosphorescence line width, which can be measured in sin
184        We describe the Madison laser-induced phosphorescence (LIP) instrument, an instrument based on
185 TP), emits a highly resolved low-temperature phosphorescence (LTP) spectrum and has the narrowest ODM
186  minutes of eye closure, using a time-domain phosphorescence measurement system.
187 the effect of scavengers, the chlorothalonil phosphorescence measurement, and varying irradiation con
188                             Frequency-domain phosphorescence measurements of the rotational dynamics
189                         Electrochemistry and phosphorescence measurements of this complex indicate a
190  detection of analytes through time-resolved phosphorescence measurements.
191 bulins have been quantified by time-resolved phosphorescence measurements.
192                       Near-infrared confocal phosphorescence microscopy was used to demonstrate the a
193  fluorescence microscopy) or P(iO(2)) (n= 7; phosphorescence microscopy) was measured continuously.
194 ed, with particular emphasis on the quenched-phosphorescence O2 sensing technique.
195 s based on oxygen-dependent quenching of the phosphorescence of a Pt-porphyrin complex immobilized on
196 ninvasively by oxygen-dependent quenching of phosphorescence of an injected dye that is excited by li
197 ated based on the quenching by oxygen of the phosphorescence of an intravenously injected palladium p
198                                  The (3)MLCT phosphorescence of each of the three coordination polyme
199 is study used steady-state and time-resolved phosphorescence of erythrosin B to monitor mobility in t
200                   The quenching mechanism of phosphorescence of Mn-doped ZnS QDs by IDA is a combined
201 tion of [1PtOEP] leads to an increase in the phosphorescence of PtOEP under ambient conditions.
202             In this study, the time-resolved phosphorescence of singlet oxygen produced by the sensit
203                         The sensitization of phosphorescence of Tb3+ bound to factor VIII subunits wa
204                           Mn2+ inhibited the phosphorescence of Tb3+ bound to HC and LC, as well as t
205 sion observed for 2 and 3 corresponds to the phosphorescence of the aromatic substrate and suggests t
206                            Consistently, the phosphorescence of the benzophenone units and the fluore
207    Upon Mg(2+) complexation in THF solution, phosphorescence of the hexathiobenzene core is turned on
208 th the purpose of amplifying the 2PA induced phosphorescence of the metalloporphyrin.
209  can be measured in human subjects using the phosphorescence of the porphyrin-protein complex bound t
210                                              Phosphorescence of tryptophan can be seen from the prote
211 oxygen-sensitive molecular probe to generate phosphorescence optical section images.
212 , photostable fluorescence, oxygen-sensitive phosphorescence or dual emission for ratiometric sensing
213  of energy transfer via an emissive process (phosphorescence) or a nonemissive process (triplet-tripl
214  Delta E(0,0) is the shift of the tryptophan phosphorescence origin, provides a measure of aromatic s
215 orm are desirable, but when fluorescence and phosphorescence originate from the same dye, it can be c
216 -temperature dual emission, fluorescence and phosphorescence, originating mainly from (1)MLCT and (3)
217 tional aspects for the estimation of various phosphorescence parameters, like intensity, radiative ra
218  TADF path (62%) and one via the T1 state as phosphorescence path (38%).
219                   Qualitative aspects of the phosphorescence phenomenon are discussed in terms of con
220            The solid-state quantum yields of phosphorescence (Phi) vary from 0.1% (1a) to 25% (1d), d
221 on spontaneous emission, and singlet-triplet phosphorescence processes--can occur on very short time
222                          A sharp increase in phosphorescence quantum efficiency is observed in a vari
223 hibit rich photophysics combined with a high phosphorescence quantum yield, are employed in red and n
224 ovements include significant increase in the phosphorescence quantum yield, higher efficiency of the
225 d organic phosphors to achieve a bright 7.5% phosphorescence quantum yield.
226 Photophysical properties (2PA brightness and phosphorescence quantum yields and lifetimes) of the new
227                         Triplet energies and phosphorescence quantum yields as well as quantum effici
228                              Remarkably, the phosphorescence quantum yields of Pd and Pt TBPs reach a
229 te Cu(I)-Au(I) covalent bonds and near-unity phosphorescence quantum yields.
230                                              Phosphorescence quenching and viscometric titrations ind
231 consumption impose limitations on the use of phosphorescence quenching in thick tissues.
232    Among existing pO(2) measurement methods, phosphorescence quenching is optimally suited for the ta
233                                  We used the phosphorescence quenching method and a specially designe
234       Tissue pO(2) and pH were determined by phosphorescence quenching microscopy and ratio imaging m
235                                              Phosphorescence quenching microscopy provided PO(2) meas
236 eric arterioles in vivo by using noninvasive phosphorescence quenching microscopy.
237 ograft, using fluorescence ratio imaging and phosphorescence quenching microscopy.
238                                              Phosphorescence quenching of certain metalloporphyrins i
239                  PO2 was measured, using the phosphorescence quenching of Pd-meso-tetra-(4-carboxyphe
240 eal oxygen consumption is measured using the phosphorescence quenching of Pd-meso-tetra-(4-carboxyphe
241 resolution can be made possible by combining phosphorescence quenching technique with multiphoton las
242                                   We adapted phosphorescence quenching techniques to determine the.Q(
243         To test this hypothesis, we utilized phosphorescence quenching techniques to measure mean mic
244                Radiolabelled microsphere and phosphorescence quenching techniques were used to measur
245 We measured thermal activation of tryptophan phosphorescence quenching to explore millisecond-range p
246 O2 uptake (VO2) ratio) profile (assessed via phosphorescence quenching) compared to muscle of primari
247  flow (radiolabelled microspheres), PO(2)mv (phosphorescence quenching), and V(O(2)) (Fick calculatio
248  decline in diaphragm microvascular PO2 (via phosphorescence quenching).
249 e nucleoprotein complex was characterized by phosphorescence, radiative decay lifetimes, and low-temp
250        Addition of poly(I) to p10 leads to a phosphorescence red shift, reduction in the zero-field s
251                     (5-BrdU)7dT yields large phosphorescence red shifts in p10 and p10-ZF, and reduct
252 tercalation with [d(CGACGTCG)](2) produces a phosphorescence redshift, while groove binding with [d(G
253 te was determined from the dependence of the phosphorescence relaxation rate on dye concentration in
254          As a general property of molecules, phosphorescence represents a cornerstone problem of chem
255 rely organic materials with room-temperature phosphorescence (RTP) are currently under intense invest
256         Although persistent room-temperature phosphorescence (RTP) emission has been observed for a f
257 anic materials that exhibit room-temperature phosphorescence (RTP) is a very attractive topic owing t
258 printed polymer (MIP)-based room temperature phosphorescence (RTP) probe by combining the RTP of Mn-d
259 ohybrids were used as novel room temperature phosphorescence (RTP) sensor to detect double stranded d
260 ion and thus enables bright room-temperature phosphorescence (RTP) with quantum yields reaching 24%.
261 er work to be the source of room-temperature phosphorescence (RTP), emits a highly resolved low-tempe
262          We devised a novel room-temperature-phosphorescence (RTP)-based oxygen detection platform by
263     In this study, a facile room-temperature phosphorescence sensor is developed to detect DA based o
264 l characteristics allowed differentiation of phosphorescence signals from the retinal vasculature and
265 ips can be obtained between the solid-matrix phosphorescence (SMP) and the percent modification of be
266 ing correlation between the magnitude of the phosphorescence spectral shift, Deltanu(0-0), and the gu
267                                              Phosphorescence spectroscopy of mixed aggregate/gels con
268 me activation, we have used frequency-domain phosphorescence spectroscopy to measure the rotational d
269            Thus, we can use frequency-domain phosphorescence spectroscopy to measure the rotational d
270 R and optical (absorption, fluorescence, and phosphorescence) spectroscopy.
271                                    (iii) The phosphorescence spectrum of Trp-151 is red-shifted in th
272 ombination of Solid Surface-Room Temperature Phosphorescence (SS-RTP) and nanotechnology has led to a
273 bre mat) with Solid Surface-Room Temperature Phosphorescence (SS-RTP) measurement for the determinati
274       The interpretation of room temperature phosphorescence studies of proteins requires an understa
275 rk is put in the context of room temperature phosphorescence studies of proteins.
276 resolved absorption, fluorescence, and (1)O2 phosphorescence studies together with fluorescence and (
277 tandards and corroborated by low-temperature phosphorescence studies, established cooperative assembl
278 efficient intersystem-crossing S1 --> Tn and phosphorescence T1 --> S0.
279 lymeric structure, sensitization of the core phosphorescence takes place with >90% efficiency.
280 reate purely organic materials demonstrating phosphorescence that can be turned on by incorporating h
281     We furthermore discuss modern studies of phosphorescence that cover topics of applied relevance,
282            The transduction of the sensor is phosphorescence; the covalently immobilized tryptamine i
283 ata formats in both the fluorescence and the phosphorescence time domains.
284 y water molecules produced unique reversible phosphorescence-to-fluorescence switching behavior.
285 s, spin-orbit coupling is less efficient and phosphorescence usually cannot compete with radiationles
286           Combined with exceptionally bright phosphorescence (varphiphos = 0.45), strong 2PA makes Pt
287 arkable ratiometric changes of intensity for phosphorescence versus fluorescence that are excitation
288 tion, and a relatively strong and long-lived phosphorescence was observed in low-temperature glasses
289  for twisted systems unexpectedly long-lived phosphorescence was observed.
290           By using single tryptophan protein phosphorescence, we follow site-specific internal protei
291 hin a cylindrical capsule gives bright green phosphorescence, while irradiation of benzil and dimetho
292 ng ocular fluorometer was designed to excite phosphorescence with a brief flash of light and to measu
293 und exhibits an additional highly structured phosphorescence with a vibronic structure corresponding
294                   Finally, the AuNC displays phosphorescence with large Stokes shift and microsecond
295  phosphorescent nanoparticles exhibit strong phosphorescence with long lifetime and large Stoke shift
296         However, previous attempts to couple phosphorescence with two-photon laser scanning microscop
297               The latter display light-green phosphorescence with unusually long lifetimes and circul
298 species (fluorescence, chemiluminescence and phosphorescence) within a few hundred nanometers from th
299                                          The phosphorescence yield for protein and model indole compo
300 ere is not a consistent effect on triplet or phosphorescence yields.

WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。
 
Page Top