戻る
「早戻しボタン」を押すと検索画面に戻ります。

今後説明を表示しない

[OK]

コーパス検索結果 (1語後でソート)

通し番号をクリックするとPubMedの該当ページを表示します
1 n the commercially available and a lab-built sprayer.
2 y than when it was delivered intranasally by sprayer.
3 ample sprayer is a conventional Z-spray type sprayer.
4 llows switching between reference and sample sprayer.
5 rate of 20-40 nl/min using an ultra-low flow sprayer.
6 g for the slight variation in signal between sprayers.
7 ously moving the sample relative to the DESI sprayer and the inlet of the mass spectrometer.
8 ystem that is physically located between the sprayer and the orifice of a mass spectrometer can serve
9  ranges from 80.7 to 94.5% between the three sprayers and is 86.5% overall.
10 ability ranges from 1 to 224% for individual sprayers and peaks of different spectral abundance.
11                          Airbrush, automatic sprayer, and sublimation matrix application methods were
12 a sections are analyzed with three lab-built sprayers, and classification accuracy for adenocarcinoma
13             The ions produced by the analyte sprayer are continuously sampled, as opposed to time-sha
14 nce neither the voltage nor positions of the sprayers are changed, the nanospray has greater spray st
15                                          The sprayers are independently controlled and do not exhibit
16 eal-time the extracted ion current from each sprayer channel.
17 ncluded changing the electrospray ionization sprayer configuration, increasing the sample injection v
18                         Small differences in sprayer construction influence the operating conditions
19                                   A multiple sprayer DESI source capable of analyzing a larger sample
20                 In this study, the impact of sprayer design and geometry on performance in desorption
21                           A miniaturized ion sprayer device is described which is suitable for coupli
22   The results demonstrated that the dual-ESI-sprayer, dual-inlet design provides reference peaks on e
23 al-ESI-sprayer, Y-shaped inlet; and dual-ESI-sprayer, dual-inlet.
24 son between the mass accuracy using dual-ESI-sprayer, dual-nozzle TOF-MS and that obtained using a do
25 r molecular-formula confirmation, a dual-ESI-sprayer, dual-nozzle version of this design was used.
26 r cone (in-spray supercharging) using a dual-sprayer ESI microchip.
27 itivity porous electrospray ionization (ESI) sprayer for the proteomic analysis of a moderately compl
28 ity to lower noise levels with the porous CE sprayer, illustrated by better signal-to-noise ratios of
29  to maximize the conduction of ions from the sprayer into the mass spectrometer.
30                                With the dual-sprayer ion source, both sprays are orthogonal to each o
31 mized for low flow rates, whereas the sample sprayer is a conventional Z-spray type sprayer.
32                                The reference sprayer is optimized for low flow rates, whereas the sam
33 s spectrometry (DESI-MS) is assessed, as the sprayer is thought to be a major source of variability.
34 ass spectrometry (CE-MS), using a porous tip sprayer, is proposed for the first time for highly sensi
35                      A new electrospray dual sprayer, LockSpray, was developed for accurate mass meas
36                                The automatic sprayer method also showed more reproducible results and
37     When using CHCA, the optimized automatic sprayer method and humidified sublimation method resulte
38 ng existing delivery systems (drenches, line sprayers, on-line dips) can significantly reduce fruit d
39                                      Between sprayers, repeatability is 16%, 9%, 23%, and 34% for hig
40       The presented results confirm that the sprayer setup needs to be closely controlled to obtain r
41 ontrolled to obtain reliable data, and a new sprayer setup with a fixed solvent capillary geometry sh
42 were compared with those obtained using dual-sprayer, single-nozzle TOF-MS.
43 le-design combinations, including single-ESI-sprayer, single-nozzle; dual-ESI-sprayer, single-nozzle;
44  single-ESI-sprayer, single-nozzle; dual-ESI-sprayer, single-nozzle; dual-ESI-sprayer, Y-shaped inlet
45 with fixed spatial relationships between the sprayer, surface, and sampling capillary.
46 ed method developed for the automatic matrix sprayer system resulted in approximately double the numb
47 e of 100 gai/ha in 200 L water using a track sprayer system, residues of these fungicides on the leaf
48                               Typically, the sprayer tip must be very close to the entrance orifice o
49 hat was centered on a point on-axis from the sprayer tip to the surface.
50 repeatability is compared for five lab-built sprayers using pork liver sections.
51                      To assess the impact of sprayer variability on tissue classification using multi
52                           When an electronic sprayer was used for matrix coating and with a high-pH (
53 ly reported proteomic application of the ESI sprayer, we evaluated preconcentration with SPME and mul
54                           Experiments with a sprayer with a fixed solvent capillary position show tha
55 e; dual-ESI-sprayer, single-nozzle; dual-ESI-sprayer, Y-shaped inlet; and dual-ESI-sprayer, dual-inle

WebLSDに未収録の専門用語(用法)は "新規対訳" から投稿できます。