戻る
「早戻しボタン」を押すと検索画面に戻ります。 [閉じる]

コーパス検索結果 (1語後でソート)

通し番号をクリックするとPubMedの該当ページを表示します
1 ich a hydrodynamic pressure is added to both electroosmotic and electrophoretic contributions is prop
2 hin a sample well generated through combined electroosmotic and hydrodynamic flows.
3 e have observed in sinusoidally oscillating, electroosmotic channel flows.
4                                          The electroosmotic component was distinguished from the diff
5 carbon nanotube demonstrates oscillations in electroosmotic current through its interior at specific
6 r capillary isoelectric focusing (CIEF) with electroosmotic displacement.
7  single fitting parameter for each molecule (electroosmotic drag coefficient).
8 certain advantages over designs that utilize electroosmotic driven flow has been fabricated and teste
9 closer to the high pH (cathode) end when the electroosmotic effect dominates.
10 smotic permeability for water as well as its electroosmotic effect, and characterized the permeabilit
11 constants for betaCD were consistent with an electroosmotic effect.
12  this pipettor consists of a microfabricated electroosmotic (EO) flow pump, a polyacrylamide groundin
13                    The surface chemistry and electroosmotic (EO) mobility of polymer microchannels la
14       We also compare the performance of our electroosmotic (EO)-driven HPLC with Agilent 1200 HPLC;
15 uidic chips using soft lithography, unstable electroosmotic flow (EOF) and cathodic drift are signifi
16 (ethylene glycol) diacrylate (PEGDA) induced electroosmotic flow (EOF) and increased the separation t
17                              Measurements of electroosmotic flow (EOF) and separation efficiency duri
18 ative standard deviation (RSD) values of the electroosmotic flow (EOF) and the first peak ((R)-(+)-BN
19  the DNA translocation relies on the induced electroosmotic flow (EOF) and the particle-nanopore elec
20  electrophoretic velocity is balanced by the electroosmotic flow (EOF) and where the sample concentra
21 port an experimental investigation of radial electroosmotic flow (EOF) as an effective means for cont
22 acking and can produce a strong and constant electroosmotic flow (EOF) at low pH.
23 at pH 9.0 for the two analytes, although the electroosmotic flow (EOF) at pH 9.0 provides sufficient
24 high as 5000 fold with an original symmetric electroosmotic flow (EOF) condition.
25                                              Electroosmotic flow (EOF) for adsorbent and exchanger pa
26                                  The annular electroosmotic flow (EOF) generated by the PEI coating a
27                             It is shown that electroosmotic flow (EOF) has much more influence on the
28 e have successfully measured the risetime of electroosmotic flow (EOF) in a microcapillary using rece
29                                          The electroosmotic flow (EOF) in a poly(dimethylsiloxane) (P
30 eby providing the relatively strong reversed electroosmotic flow (EOF) in capillary electrochromatogr
31 e the extent of intraparticle, or perfusive, electroosmotic flow (EOF) in CEC capillaries packed with
32 k is an analytical and experimental study of electroosmotic flow (EOF) in cylindrical capillaries wit
33        In comparison to glass microchannels, electroosmotic flow (EOF) in native PC channels is low a
34 pillary surface responsible for the reversed electroosmotic flow (EOF) in the columns during CEC oper
35                                              Electroosmotic flow (EOF) is commonly utilized in microf
36                                              Electroosmotic flow (EOF) is induced as the driving forc
37                           A reduction in the electroosmotic flow (EOF) is often desirable in glass mi
38                                              Electroosmotic flow (EOF) is used to enhance the deliver
39 was performed with the use of thiourea as an electroosmotic flow (EOF) marker.
40 with DNA electrophoresis where a substantial electroosmotic flow (EOF) may be detrimental to the sepa
41                                              Electroosmotic flow (EOF) measurements in modified and u
42                                              Electroosmotic flow (EOF) or electro-osmosis has been sh
43 n the electric field eliminated the need for electroosmotic flow (EOF) or hydrodynamic flow for dropl
44             A flow-based interface that uses electroosmotic flow (EOF) provides continuous injection
45 strate here a new electrokinetic phenomenon, Electroosmotic flow (EOF) rectification, in synthetic me
46             Native TPE microchannels support electroosmotic flow (EOF) toward the cathode, with an av
47 responding opposition of electrophoretic and electroosmotic flow (EOF) velocities.
48                                              Electroosmotic flow (EOF) was driven across the CNMs by
49                                              Electroosmotic flow (EOF) was monitored in glass microfl
50 e channels to control analyte adsorption and electroosmotic flow (EOF) while maintaining separation e
51 fords monolithic CEC columns that facilitate electroosmotic flow (EOF) while preventing ionized analy
52  dynamic coating method that provided stable electroosmotic flow (EOF) with respect to pH.
53                                              Electroosmotic flow (EOF) with two or more fluids is oft
54                                              Electroosmotic flow (EOF) within channels was used to de
55  ethanol (NPE), which is only transported by electroosmotic flow (EOF), a positive correlation betwee
56 e microchannel walls enables reversal of the electroosmotic flow (EOF), enabling cations, instead of
57 reason for this asymmetry, we identified the electroosmotic flow (EOF), which is the water transport
58                                           An electroosmotic flow (EOF)-based pump, integrated with a
59 ization methods: chemical, hydrodynamic, and electroosmotic flow (EOF)-driven mobilization.
60 e to suppress analyte adsorption and control electroosmotic flow (EOF).
61  and the ability to control the magnitude of electroosmotic flow (EOF).
62 reagents were transported into the system by electroosmotic flow (EOF).
63     Flow through these filters was driven by electroosmotic flow (EOF).
64 d and overcome by the shear force induced by electroosmotic flow (EOF, i.e. the water flow over surfa
65 urfactants on PDMS was studied by performing electroosmotic flow (microEOF) measurements.
66 ppears to require a dilute polymer solution, electroosmotic flow (preferably countercurrent to the di
67 the SDS micelles velocity is faster than the electroosmotic flow (using acidic buffer), MCDS was cond
68 eld acting on their charge and (ii) the bulk electroosmotic flow (which is directed toward the cathod
69                Injecting neutral analytes by electroosmotic flow affords a 10-fold or greater decreas
70 rt of charged species in pressure-driven and electroosmotic flow along nanoscale channels having an e
71             On the negative side, nonuniform electroosmotic flow along the capillary or microfluidic
72 rescent molecules through the tissue by both electroosmotic flow and electrophoresis.
73 ects of increasing the sample plug length on electroosmotic flow and electrophoretic current agreed q
74  allows us to work under conditions in which electroosmotic flow and electrophoretic forces add or op
75 uded poly(vinylpyrrolidone) to eliminate the electroosmotic flow and mannitol to enhance the separati
76    The perfusate was continuously sampled by electroosmotic flow and mixed online with Cy5-labeled in
77 ied surfaces exhibited substantially reduced electroosmotic flow and nonspecific adsorption of protei
78 st time the independent optimization of both electroosmotic flow and retention properties in CEC colu
79 ify and distinguish the contributions of the electroosmotic flow and the electrophoretic force on tra
80 ns describing the generation of vorticity in electroosmotic flow are derived using a wall-local, stre
81  capillary-to-capillary reproducibilities of electroosmotic flow are very good with relative standard
82 lish a pH gradient as well as to control the electroosmotic flow arising from the use of uncoated fus
83 rthermore, application of UV modification to electroosmotic flow around a 90 degrees turn results in
84  needed to flush the PPMs since they support electroosmotic flow as cast.
85 urb the capillary electrophoresis separation electroosmotic flow as evidenced by the observation that
86                                      (2) The electroosmotic flow at reversed polarity (negative) mode
87 capillary, with neutral analytes injected by electroosmotic flow at up to 1 order of magnitude faster
88                                              Electroosmotic flow between interdigitated electrodes wa
89 with an electric field prediction, to obtain electroosmotic flow bulk fluid velocity measurements.
90     Analytes are injected at the velocity of electroosmotic flow but are retained at the interface of
91 chain and mixed PEG-silane coatings suppress electroosmotic flow by more than 90%, whereas the short-
92 ty of the polymers, and the direction of the electroosmotic flow can be altered without degrading chr
93 n optical force was applied to an orthogonal electroosmotic flow carrying a hydrodynamically pinched,
94                                              Electroosmotic flow changes on the order of 100% (1.6-3.
95                                 The reversed electroosmotic flow could be explained by slower depleti
96 ography, neutral analytes can be injected by electroosmotic flow directly from a sample matrix into a
97 cal description of band broadening caused by electroosmotic flow dispersion (EOFD) and the experiment
98 external electric field, creating nonuniform electroosmotic flow distributions.
99       This research proposes a point-of-care electroosmotic flow driven microfluidic device for rapid
100                                   Changes in electroosmotic flow during sample stacking and separatio
101                                              Electroosmotic flow dynamics during a field-amplified sa
102 article velocity due to convection driven by electroosmotic flow exceeded that of electrophoresis at
103  the temperature increase in the presence of electroosmotic flow for a 100-, 200-, and 300-microm cha
104 ngle zone (peak) which is separated from the electroosmotic flow front and any other interfering mole
105                                              Electroosmotic flow further modulates the local field gr
106                                              Electroosmotic flow has been monitored in a capillary us
107 The CE separation was performed at near-zero electroosmotic flow in a capillary with neutral, hydroph
108 onitoring technique for measuring an average electroosmotic flow in a microfluidic device with a cros
109 ocity, and late-time solute distribution for electroosmotic flow in a tube and channel at zeta potent
110                         It is found that the electroosmotic flow in aminated PMMA microchannels is re
111  formamide, which has been shown to diminish electroosmotic flow in glass microchannels by over 5 ord
112 trated by our laboratory to nearly eliminate electroosmotic flow in glass microchannels was employed
113 redictability and constancy over time of the electroosmotic flow in microchannels is an important con
114   Protocols are described for control of the electroosmotic flow in microfabricated channels in Vivak
115 ar, this study investigates perturbations of electroosmotic flow in open capillaries that are due to
116                        We have characterized electroosmotic flow in plastic microchannels using video
117  In this paper, the Taylor dispersion due to electroosmotic flow in such a partially coated capillary
118                                 In addition, electroosmotic flow in the device plays a critical role
119 orescence detector, to determine the rate of electroosmotic flow in the entire capillary.
120 dsorption of the virus capsids, and suppress electroosmotic flow in the pore.
121 ith peptides and proteins and to reverse the electroosmotic flow in the separation channel.
122 ld resulting from the iontophoretic current, electroosmotic flow in the tissue would carry solutes co
123                                          The electroosmotic flow increased from 4.1 x 10(-4) to 6.8 x
124             Rectification of ion current and electroosmotic flow increased with increasing electric f
125  obtained for peak height and peak area with electroosmotic flow injection is comparable to that obta
126 netic stacking of neutral analytes utilizing electroosmotic flow is demonstrated with discontinuous (
127                                              Electroosmotic flow is fluid motion driven by an electri
128 sfer across the pore/solution interface when electroosmotic flow is operative.
129       Simple dimensional arguments show that electroosmotic flow is potentially as important as diffu
130                               Suppression of electroosmotic flow is proposed as a means of reducing m
131        Moreover, these findings suggest that electroosmotic flow is significant for particle transpor
132 of separation selectivity and the normalized electroosmotic flow mobility.
133 t disk UMEs in low ionic strength solutions, electroosmotic flow occurring at the glass insulation of
134 ution velocity to elucidate the influence of electroosmotic flow on transport of bacteria near the el
135                          Manipulation of the electroosmotic flow opens the door to hydrodynamic modul
136       Ultrathin pnc-Si membranes enable high electroosmotic flow per unit voltage.
137                            A mismatch in the electroosmotic flow rate at this junction led to the gen
138 icrostructure of the rough microchannel, the electroosmotic flow rate decreases with the Debye length
139                                          The electroosmotic flow rate increases with the roughness nu
140 e number density of roughness are given, the electroosmotic flow rate is enhanced by the increase of
141 tection, while simultaneously monitoring the electroosmotic flow rate.
142 ionally in such systems due to a mismatch in electroosmotic flow rates or hydrostatic pressure differ
143 oxidation or silanization, can influence the electroosmotic flow rates through pnc-Si membranes by al
144 age-biased OmpF nanopore is dominated by the electroosmotic flow rather than the electrophoretic forc
145 cation by current-voltage (I-V) response and electroosmotic flow rectification by transport of a zwit
146  not only ion current rectification but also electroosmotic flow rectification.
147                       In all cases, however, electroosmotic flow resulted in significantly less sampl
148               We propose that this is due to electroosmotic flow separation, a high-salt electrokinet
149                In the symmetric case for the electroosmotic flow so induced, the velocity field and t
150 ion exchange beads, which produce convergent electroosmotic flow streams.
151 ingle-column ITP configuration together with electroosmotic flow suppression and high leading ion con
152  of dynamic wall coatings for the purpose of electroosmotic flow suppression can have a significant i
153 d performance associated with the use of the electroosmotic flow switching system in a point-of-care
154 otypic hippocampal slice cultures (OHSCs) by electroosmotic flow through an 11 cm (length) x 50 mum (
155 surface charge, we can control the amount of electroosmotic flow through the nanopore, which affects
156          The experimental inquiry focuses on electroosmotic flow under a uniform applied field in cap
157 ontrol (FEFC) modifies the zeta potential of electroosmotic flow using a transverse electric field ap
158  be easily modified to control inertness and electroosmotic flow using a variety of chemical procedur
159  detail the means of achieving bidirectional electroosmotic flow using an array of alternating curren
160  provides a mechanism for fine tuning of the electroosmotic flow velocity when 2-acrylamido-2-methyl-
161 xhibits a lower ion conductance and a higher electroosmotic flow velocity, whereas, in the tip-to-bas
162 o cause a small, but definite, change in the electroosmotic flow velocity.
163 nel has a higher ion conductance and a lower electroosmotic flow velocity.
164 ermined by several parameters, including the electroosmotic flow velocity.
165  interplay of concentration polarization and electroosmotic flow with respect to the observed concent
166      By using high concentrations of buffer, electroosmotic flow within uncoated channels of a microf
167 e control, adjustment, and modulation of the electroosmotic flow without using wall coatings or chang
168 or zone narrowing to occur assume negligible electroosmotic flow, a relatively constant electric fiel
169 the solvent used to cast the polymer enables electroosmotic flow, allowing the separation channel to
170                                              Electroosmotic flow, and the resulting transport of neur
171 f the applied potential and the direction of electroosmotic flow, either anions or cations can be con
172 expensive, and, in addition to reversing the electroosmotic flow, provides excellent separation effic
173 pillary electrophoretic buffer modulated the electroosmotic flow, resulting in optimum separation of
174  Our model yields analytical expressions for electroosmotic flow, species transport velocity, streamw
175 horesis channel and a portion is injected by electroosmotic flow, termed the "discrete injector".
176 ause of the simplicity and rapid response of electroosmotic flow, this technique may be useful for ne
177 ularly the high voltage used for driving the electroosmotic flow, upon the background current, potent
178 ged and neutral glycans, such as influencing electroosmotic flow, using complexation/interaction base
179 f 0.5% v/v, which effectively suppresses the electroosmotic flow, was added to the background electro
180   As the effect occurs within an oscillating electroosmotic flow, we refer to it here as an electroki
181  Through simulations, we found that reversed electroosmotic flow, which filled the pore with aqueous
182 nique in which the mobile phase is driven by electroosmotic flow, while the sorbent layer is pressuri
183                The mobile phase is driven by electroosmotic flow, while the system is pressurized in
184 pike-shaped transients was convection due to electroosmotic flow.
185 a second via a low-voltage pulse that drives electroosmotic flow.
186 s experienced transport by convection due to electroosmotic flow.
187 cies are driven along them in the absence of electroosmotic flow.
188 e relative standard deviations (RSDs) of the electroosmotic flow.
189 d mixing of two confluent streams undergoing electroosmotic flow.
190 d the outlet end) by hydrodynamic flow or by electroosmotic flow.
191  for species transport by electrophoresis or electroosmotic flow.
192 ss the pore under different biases caused by electroosmotic flow.
193 fficiency from the flow profile generated by electroosmotic flow.
194  ODS columns are characterized by switchable electroosmotic flow.
195 sity, and self-coating property for reducing electroosmotic flow.
196 of a 100-pL mixer for liquids transported by electroosmotic flow.
197 olarities and the capabilities of a reversed electroosmotic flow.
198 ethylammonium hydroxide, for reversal of the electroosmotic flow.
199 oupling to all other mobile ions, causing an electroosmotic flow.
200 mobilities of free protein, free ligand, and electroosmotic flow.
201 ly affected the elution times by varying the electroosmotic flow.
202  with a neutral coating exhibiting near-zero electroosmotic flow.
203          The use of acidic ES suppressed the electroosmotic flow; allowing the electrokinetic movemen
204  between the electric and velocity fields in electroosmotic flows are discussed.
205 flows, the bulk flow region of time periodic electroosmotic flows are rotational when the diffusion l
206 sional, time-independent, and time-dependent electroosmotic flows driven by a uniform electric field
207                                 In contrast, electroosmotic flows generally yield identical speeds fo
208        Analytical solutions of time periodic electroosmotic flows in two-dimensional straight channel
209 us drag forces on deflecting microtubules in electroosmotic flows were studied theoretically and expe
210 t is also shown that, unlike the steady pure electroosmotic flows, the bulk flow region of time perio
211 ence plays an important role in microchannel electroosmotic flows.
212 trolyte concentrations induce convection via electroosmotic flows.
213 ssure on the velocity and vorticity field of electroosmotic flows.
214 mide (CTAB) is shown to provide reproducible electroosmotic flows.
215  the presence of H2O2 is capable of inducing electroosmotic fluid flow that can be switched on and of
216 charges in the extracellular space to create electroosmotic fluid flow within the extracellular space
217                                          The electroosmotic fluid velocity is used to analyze late-ti
218 t first reported by Anderson and Idol on the electroosmotic flux in capillaries with axial variations
219  position of streptavidin in electrophoretic-electroosmotic focusing (EEF) experiments was monitored
220                           An electrophoretic-electroosmotic focusing (EEF) method was developed and u
221  fluid velocity through the pore at constant electroosmotic force is determined by fitting the theore
222 rge at salt concentrations below 5 mM, where electroosmotic forces are more significant.
223        EEF uses opposing electrophoretic and electroosmotic forces to focus and separate proteins and
224  proteins, responding to electrophoretic and electroosmotic forces, have long been proposed as the se
225 ions are due to the sum of iontophoretic and electroosmotic forces.
226 ated in the capillary by electrophoretic and electroosmotic forces.
227                                              Electroosmotic manipulation of fluids was demonstrated u
228                                      Initial electroosmotic mobilities (EOM) of (8.3+/-0.2)x10(-4) cm
229 rits are formed in open capillaries, and the electroosmotic mobilities are calculated and compared to
230                             The experimental electroosmotic mobilities compare quantitatively to mobi
231 confinement within the channels, the average electroosmotic mobilities decrease.
232                                  We measured electroosmotic mobilities in NaCl solutions from 0.1 to
233                 We report the measurement of electroosmotic mobilities in nanofluidic channels with r
234               At kappah approximately 4, the electroosmotic mobilities in the 27, 54, and 108 nm chan
235 he Smoluchowski equation accurately predicts electroosmotic mobilities in the nanochannels.
236 icrochannels with grafted surfaces exhibited electroosmotic mobilities intermediate to those displaye
237                             The differential electroosmotic mobilities of the enzyme and substrate, L
238 ied with PEMs, they demonstrate very similar electroosmotic mobilities.
239 d separations and exhibited little change in electroosmotic mobility between pH 2.8 and pH 7.5.
240 site-binding model, we demonstrated that the electroosmotic mobility could be controlled qualitativel
241 measurement of the separation efficiency and electroosmotic mobility for multiple microfluidic device
242 e experimental observations of (i) a maximum electroosmotic mobility for the first scenario as the pH
243  (ii) the inversion and maximum value of the electroosmotic mobility for the second scenario when the
244 ed with 0.5 wt % undecylenic acid (UDA), the electroosmotic mobility in a modified PDMS channel rises
245                 We report on measurements of electroosmotic mobility in polymer microchannels and sil
246 hannel (kappah = 1) is 5-fold lower than the electroosmotic mobility in the 2.5 mum channel (kappah =
247 nels exhibit maxima, and at 0.1 mM NaCl, the electroosmotic mobility in the 27 nm channel (kappah = 1
248 s for controlling the flow direction and the electroosmotic mobility in the channels.
249 ciprocal of the solvent viscosity, while the electroosmotic mobility increases in a linear fashion wi
250 lution rely on either streaming potential or electroosmotic mobility measurement techniques, both of
251 l with a dynamic coating of DDM generates an electroosmotic mobility of (5.01 +/- 0.09) x 10(-4) cm(2
252 ow (EOF) toward the cathode, with an average electroosmotic mobility of 1.3 x 10(-4) cm(2) V(-1) s(-1
253 laries treated with this coating produced an electroosmotic mobility of 2.8 +/- 0.2 x 10(-6) cm(2).V(
254                                   Unlike the electroosmotic mobility of oxidized PDMS, the electroosm
255 lectroosmotic mobility of oxidized PDMS, the electroosmotic mobility of the grafted surfaces remained
256 ctrophoresis, it is important to control the electroosmotic mobility of the running buffer and the fa
257                                 The very low electroosmotic mobility results in a 200 min separation
258          The general equation to predict the electroosmotic mobility suggested here also indicates th
259                                          The electroosmotic mobility was stable in response to air ex
260 uggested here also indicates the increase of electroosmotic mobility with temperature.
261                         It is shown that the electroosmotic mobility, induced by an electric field ap
262 gate species transport by electrophoretic or electroosmotic motion in the curved geometry of a two-di
263  substrate or substrate plus inhibitor using electroosmotic perfusion (EOP).
264 e have developed an approach that integrates electroosmotic perfusion of tissue with a substrate-cont
265                                          The electroosmotic perfusion-microdialysis probe and associa
266 s and define three new concepts based on the electroosmotic potential distribution.
267 t, wall shear stress, and vorticity in mixed electroosmotic/pressure driven flows are presented for t
268 sign allows for the production of a fritless electroosmotic pump and easy replacement of the ion exch
269                                Moreover, the electroosmotic pump can generate high flow rates over an
270   Here, we construct an open-channel on-chip electroosmotic pump capable of generating pressures up t
271                                           An electroosmotic pump is incorporated at the end of the el
272                                   A fritless electroosmotic pump with reduced pH dependence has been
273 d with polyacrylamide for the elimination of electroosmotic pumping and protein adsorption onto the c
274        A multichannel architecture that uses electroosmotic pumping principles provides the necessary
275                      The micropump, based on electroosmotic pumping principles, has a multiple open-c
276                                The different electroosmotic pumping rates formed by local zeta potent
277 w splitting, capillary electrophoresis (CE), electroosmotic pumping, and electrospray ionization (ESI
278 y in a wide range of applications, including electroosmotic pumping, liquid chromatographic separatio
279 er capillary and samples were transferred by electroosmotic pumping.
280                            We have developed electroosmotic pumps (EOPs) fabricated from 15-nm-thick
281                                              Electroosmotic pumps are arguably the simplest of all pu
282 ificant advantage over previously fabricated electroosmotic pumps, which typically have a more limite
283 ing chromatography columns, micromixers, and electroosmotic pumps.
284                                              Electroosmotic sampling is a potentially powerful method
285 approach is to define conditions under which electroosmotic sampling minimizes damage to the tissue,
286 ltz-Smoluchowski velocity is the appropriate electroosmotic slip condition even for high-frequency ex
287                           The effect of this electroosmotic solvent flow on the binding of a neutral
288 meabilization sites, and electrophoretic and electroosmotic transport by the electric pulses.
289                                              Electroosmotic transport of ascorbate occurred at a negl
290 t short distances (<100 mum), advection from electroosmotic transport of the barrel solution may sign
291 gh nanofunnels, which suggest the asymmetric electroosmotic transport stems from an induced pressure
292 dimensional simulations of ion transport and electroosmotic transport through nanofunnels, which sugg
293                            The efficiency of electroosmotic transport was also shown to be a function
294 gher than the run buffer conductivities, the electroosmotic velocities are such that there is less fl
295  near the IDZ when their electrophoretic and electroosmotic velocities balance.
296                   These results suggest that electroosmotic velocities of solute molecules are determ
297               We quantitatively describe the electroosmotic velocity component experienced by the sub
298                                          The electroosmotic velocity of the neutral molecule, acetami
299  order unity, based on channel depth and rms electroosmotic velocity.
300                        First, we measure the electroosmotic wall mobility of a borosilicate rectangul

 
Page Top